Occlusal Force (occlusal + force)

Distribution by Scientific Domains


Selected Abstracts


Influence of forces on peri-implant bone

CLINICAL ORAL IMPLANTS RESEARCH, Issue S2 2006
Flemming Isidor
Abstract: Occlusal forces affect an oral implant and the surrounding bone. According to bone physiology theories, bones carrying mechanical loads adapt their strength to the load applied on it by bone modeling/remodeling. This also applies to bone surrounding an oral implant. The response to an increased mechanical stress below a certain threshold will be a strengthening of the bone by increasing the bone density or apposition of bone. On the other hand, fatigue micro-damage resulting in bone resorption may be the result of mechanical stress beyond this threshold. In the present paper literature dealing with the relationship between forces on oral implants and the surrounding bone is reviewed. Randomized controlled as well as prospective cohorts studies were not found. Although the results are conflicting, animal experimental studies have shown that occlusal load might result in marginal bone loss around oral implants or complete loss of osseointegration. In clinical studies an association between the loading conditions and marginal bone loss around oral implants or complete loss of osseointegration has been stated, but a causative relationship has not been shown. [source]


Evaluation of minimum interdental threshold ability in dentate female temporomandibular disorder patients

JOURNAL OF ORAL REHABILITATION, Issue 5 2010
E. M. KOGAWA
Summary, Minimum interdental threshold is the smallest thickness that can be detected between teeth during an occlusion and has an influence on the occlusal force and on the control of mandibular movements. The aim of this study was to assess the possible association of the signs and symptoms of temporomandibular disorders (TMD) with the ability to detect a minimum interdental threshold. Two hundred women were equally divided into four groups: asymptomatic (control), subjects with masticatory muscle pain, with articular [temporomandibular joint (TMJ)] pain and mixed (muscular and articular pain). Evaluation of the ability to detect a minimum interdental threshold was performed using aluminium foils with 0·010, 0·024, 0·030, 0·050, 0·080 and 0·094 mm of thickness in the premolar region. A total of 20 tests with each thickness for each patient were performed, starting with the thickest foil (0·094 mm) and ending with the thinnest one. The myogenic pain and articular groups presented significantly higher threshold values (0·020 and 0·022 mm, respectively), when compared to the control. Both groups reached the level of certain perceptiveness only at 0·030 mm. No significant correlation was found between minimum interdental threshold and age. These results suggest that discrimination of thicknesses can be disturbed as a consequence of TMD manifestations and not the cause of it. Clinicians should, therefore, be aware that changes on muscles and TMJ can secondarily lead to occlusion changes. The mechanisms involved in this process, however, are not well understood and warrant further investigation. [source]


Automatic regulation of occlusal force because of hardness-change of the bite object

JOURNAL OF ORAL REHABILITATION, Issue 1 2008
A. SHIMADA
Summary, It is considered that the information of chewed food, such as size and texture, is important for smooth mastication. In this study, we analysed aspects of the control of occlusal force, by experimentally reproducing situations in which the hardness of food changed unpredictably during mastication, using a device that utilized a 3-sectioned urethane rubber piece with different hardness values. Seven healthy subjects were instructed to perform repetitive jaw open,close movements paced by a metronome (1·0 Hz) and to maintain constant occlusal force throughout the trial. Using the device, the following parameters were measured during the first to fifth strokes after changing the hardness, peak value, impulse, duration and time to peak of occlusal force in the waveform of occlusal force, cycle time of open,close jaw movements, jaw gape and maximum speed in the closing phase in the waveform of the jaw movements. Each parameter value was statistically analysed by anova with Fisher's least significant difference method (P < 0·05). Peak occlusal force, impulse and jaw gape were significantly affected by the change in hardness, while an increase in hardness caused increases in the values for peak occlusal force and impulse against the instruction, after which those values remained constant while the subjects occluded the same level of hardness. Our results indicate that the level of the resulting occlusal force is regulated automatically according to the hardness of the chewed food. We concluded that occlusal force was adapted for efficient mastication when the hardness of foods changes unpredictably. [source]


Association of tightly locked occlusion with temporomandibular disorders

JOURNAL OF ORAL REHABILITATION, Issue 3 2007
M.-Q. WANG
summary, The association between teeth loss and temporomandibular disorders (TMD) is still inconclusive. A kind of secondary changes of the occlusion after teeth lose called the tightly locked occlusion (TLO), defined as the occluding contact that delivers angled occlusal force on the drifted neighbour and/or the tipped antagonists of the lost posterior teeth, was hypothesized to be association with TMD. The study aimed at investigating the association between the TLO and TMD. A total of 113 posterior-teeth losing patients, 64 with TMD symptoms (group of TMD) and 49 without (group of TMD-Free) were included. Study casts and joint radiographs were made to diagnose the TLO and joint morphological changes. The simultaneous contribution of the potential variables of gender, age, tooth losing number, the TLO, joint symmetry and signs of osteoarthrosis shown on radiographs were tested through binary logistic regression analysis. In women, the TLO entered into logistic model, and had an effect on the incidence of TMD (P = 0·008). The odds ratio of with-TLO versus without-TLO is 2·6 (95% CI: 1·2, 5·8) after controlling for the effect of gender. Age, tooth lose number, joint asymmetry or osseous changes had no effect on the incidence of TMD. The tightly locked occlusion is associated with some signs and symptoms of TMD. Randomized controlled trials will be needed in further studies to test the hypothesis that treatment of a TLO, as defined in the present study, will have a beneficial effect on the signs and symptoms of TMD. [source]


Individual difference in the number of chewing strokes and its determinant factors

JOURNAL OF ORAL REHABILITATION, Issue 2 2006
T. UEDA
summary, This study was carried out to clarify the distribution and individual variation of the number of chewing strokes before last swallowing (NCS), as well as to assess the factors that affect NCS, when dentulous adults masticate the same type of food. NCS was measured in 75 dentulous adults using gummy candies. Measurements were repeated five times to obtain the average and variation. Moreover, the correlation of NCS with various physiological factors (lateral width and vertical distance of chewing loop, angle of opening and closing path, opening and closing time, occlusal time, masticatory performance, number of occlusal contact points, occlusal contact area, occlusal force, flow rate of whole saliva and viscosity of whole saliva) and 28 personality factors were analysed by both of the single variate analysis and the multivariate analysis. The results showed that the average NCS was 41·0 ± 16·0, and the coefficient of variation was 0·090 ± 0·040. Additionally, the correlation between NCS and each factor was investigated, and found that any of these factors did not function as a single determinant for NCS. Following, determinant factors were further examined by the stepwise method of linear multiple regression analysis. From these analysis we found that when opening and closing time and four personality factors were combined, they were significantly involved in determining NCS (P < 0·01). Therefore, we conclude that individual NCS until last swallowing is not determined by a sole, specific physiological factor but is affected by multiple factors including personality. [source]


The effect of retainer thickness on posterior resin-banded prostheses: a finite element study

JOURNAL OF ORAL REHABILITATION, Issue 11 2004
T.-S. Lin
summary, According to its design concept, a resin-bonded prosthesis, compared with the conventional fixed partial denture, is a weak and unstable structure. Therefore, a resin-bonded prosthesis induces a higher failure rate, especially in the posterior region. Recently, adhesion agents have been profoundly improved. However, the design guidelines of posterior resin-bonded prostheses (RBP) have seldom been evaluated from a biomechanical perspective. The objective of this study was to investigate the biomechanical effects of the retainer thickness on posterior RBP using the finite element method. A solid model of a posterior mandibular resin-bonded prosthesis, which employed the second molar and second premolar as the abutment teeth, was constructed and meshed with various retainer thickness (ranging from 0·2 to 1·0 mm). Horizontal and vertical loadings of 200 N were applied respectively at the central fossa of the pontic to examine the stress level at the interface between the retainer and abutment teeth. All exterior nodes in the root, below the cementoenamel junction were fixed as the boundary condition. The results showed that horizontal loading would induce higher interfacial stresses than the vertical loading which indicated that the horizontal component of the occlusal force plays a more important role in evaluating the debonding phenomenon. Further, the peak interfacial stresses increased as the retainer thickness decreased and, based on the fitted relation between retainer thickness and interfacial stresses, a 0·4 mm retainer thickness was suggested as the minimum required to prevent severe interfacial stresses increasing. [source]


Finite element analysis of the effect of the bucco-lingual position of artificial posterior teeth under occlusal force on the denture supporting bone of the edentulous patient

JOURNAL OF ORAL REHABILITATION, Issue 6 2003
G. Nishigawa
summary, To improve the quality of the complete denture prosthesis, the bucco-lingual position of the artificial posterior teeth must be determined with consideration of the shape of the maxillary and mandibular residual ridge and the relationship between them. The arrangement of posterior artificial teeth should be considered not only for the denture stability but also for the avoidance of high pressure on the supporting structures. A two-dimensional finite element method program to investigate the statics for the contour of the complete denture and the residual ridge was developed. With this program, the effect of the bucco-lingual position of the artificial posterior teeth under occlusal force on the denture supporting bone could be investigated. [source]


Relation between clenching strength and occlusal force distribution in primary dentition

JOURNAL OF ORAL REHABILITATION, Issue 3 2003
H. Karibe
summary, The aim of this study was to investigate the relation between clenching strength and occlusal force distribution in primary dentition. Twenty healthy children with normal occlusions: 11 boys and 9 girls, ages 3·2,5·8 years (avg. 4·5 years) were selected. Setting the bilateral masseter muscular activity at maximum clenching in full intercuspation as 100%, the occluding forces at 20, 40, 60, 80 and 100% clenching were recorded with pressure-sensitive sheets (Dental Prescale 50H, type R, Fuji Photo Film Co.), and the force of each primary tooth was analysed by computer (Occluzer FPD703). Occlusal force distribution was expressed as a percentage of the total occlusal force of each tooth and was compared between each clenching. There were no significant differences between various clenching strengths in the occlusal force distribution in primary dentition [one-way repeated-measures analysis of variance (anova)]. Thus, the results of the present study suggest that the distribution of occluding forces on a primary dental arch had its own pattern and that the clenching strength had no effect on that pattern. These patterns may be useful in determining occlusal function in children. [source]


Changes in occlusal force and occlusal contact area after active orthodontic treatment: a pilot study using pressure-sensitive sheets

JOURNAL OF ORAL REHABILITATION, Issue 5 2002
M. H. Sultana
The aim of this study was to investigate functional changes in occlusion during retention. Data on occlusal force (OcFr) and occlusal contact area (OcAr) was obtained using the pressure-sensitive sheet, from a treated group (20 female patients) who had had four premolar extractions and treatment with standard edgewise appliances, and a control sample who matched the treated group of retainer for sex, age and Angle classification at 1 year after removal. A repeated measures analysis of variance showed that the mean values of total OcFr and OcAr in the treatment group gradually increased during retention and were 669·3 N and 15·1 mm2, respectively, at 1 year after removal of retainer. The increases of OcFr and OcAr were larger in the molar region, especially at the second molar. At 1 year after removal of retainer, OcFr and OcAr in the second molar were significantly larger in the treatment group than in the control sample, and a similar distribution pattern of OcFr and OcAr to those in normal occlusion was seen. These results suggested that balanced OcFr and OcAr might be obtained during and after retention, due to the settling of molars that had been discluded by active orthodontic treatment. [source]


Biomechanical effects of double or wide implants for single molar replacement in the posterior mandibular region

JOURNAL OF ORAL REHABILITATION, Issue 10 2000
Y. Sato
Double implants have been thought to have biomechanical advantages for single molar replacement. To evaluate the effectiveness of double implants versus a wide implant, the vertical forces and torque on each implant were calculated by three-dimensional geometric analysis. Buccal load (100 N) perpendicular to cuspal inclination (20°) was applied at the occlusal surface of the superstructure. The three kinds of load points (A, B, C) were 1·5, 3·5, and 5·5 mm from the mesial contact point, respectively. Three implants were compared: mesial and distal double implants (, 3·3 mm), and a wide implant (, 5 mm). The wide implant showed torque around the long axis (1·8,15·0 N · cm) whereas double implants had no torque. On the other hand, the vertical forces on the mesial double implant were both smaller (60%: loaded at point C) and larger (140%: loaded at point A) than the wide implant. Given the smaller surface area of the mesial double implant, this large force may generate much higher stress in the peri-implant bone. These results suggest that the biomechanical advantage of double implants for single molar replacement is questionable when the occlusal force is loaded at the occlusal surface near the contact point. [source]


Relationship between the prognosis of periodontitis and occlusal force during the maintenance phase , a cohort study

JOURNAL OF PERIODONTAL RESEARCH, Issue 5 2010
N. Takeuchi
Takeuchi N, Ekuni D, Yamamoto T, Morita M. Relationship between the prognosis of periodontitis and occlusal force during the maintenance phase , a cohort study. J Periodont Res 2010; 45: 612,617. © 2010 John Wiley & Sons A/S Background and Objective:, Few studies have longitudinally investigated the relationship between periodontal disease progression and occlusal factors in individual subjects during the maintenance phase of periodontal therapy. The aim of this cohort study was to investigate the relationship between biting ability and the progression of periodontal disease in the maintenance phase. Material and Methods:, A total of 194 patients were monitored for 3 years during the maintenance phase of periodontal therapy. The subjects with disease progression (Progress group) were defined based on the presence of , 2 teeth demonstrating a longitudinal loss of proximal attachment of , 3 mm or tooth-loss experience as a result of periodontal disease during the study period. The subjects with high occlusal force were diagnosed as men who showed an occlusal force of more than 500 N and women who showed an occlusal force of more than 370 N. The association between biting ability and the progression of periodontitis was investigated using logistic regression analysis. Results:, There were 83 subjects in the Progress group and 111 subjects in the Non-progress group. A backward, stepwise logistic regression model showed that the progression of periodontal disease was significantly associated with the presence of one or more teeth with a high clinical attachment level (CAL) of , 7 mm (odds ratio: 2.397; 95% confidence interval: 1.306,4.399) (,p = 0.005) and low occlusal force (odds ratio: 2.352; 95% confidence interval: 1.273,4.346) (,p = 0.006). Conclusion:, The presence of one or more teeth with a high CAL of , 7 mm and low occlusal force might be possible risk factors for periodontal progression in the maintenance phase of periodontal therapy. [source]


Failure Modes with Point Loading of Three Commercially Available Denture Teeth

JOURNAL OF PROSTHODONTICS, Issue 6 2008
Andrew R. Moffitt DDS
Abstract Purpose: A common problem associated with implant-supported prostheses is the fracture of denture teeth. This study was designed to compare the fracture modes of three denture teeth by compressive load at a 30° off-axis angle. Material and Methods: Three denture teeth (Vident Duostat, Ivoclar Vivadent, and Dentsply Trubyte) processed to two denture base processing systems [injection-molded (IM) SR-Ivocap system and compression-molded (CM) denture base resin] were evaluated. Each specimen was processed to a metal framework. Ultimate failure strength of each system when point loaded at a 30° off-axis angle was recorded, along with a visual inspection of each specimen. Results: The average load fracture for each group was (in N): Vident CM 1106.97 ± 223.20, Vident IM 1168.18 ± 322.52, Dentsply CM 1098.08 ± 286.32, Dentsply IM 1023.80 ± 282.45, Ivoclar CM 1616.98 ± 204.87, and Ivoclar IM 1373.54 ± 282.58. There was a significant difference between the groups and the Ivoclar CM group. The Ivoclar CM group had the highest average load force, and the Dentsply IM group had the lowest average load force. On average, the teeth within the groups fractured at a higher compression force than the average maximum occlusal force in natural dentition. Dentsply and Vident denture teeth fractured more horizontally, and the Ivoclar denture teeth fractured more vertically within the groups. There was no significant difference among the groups between the IM and CM processing methods. Conclusions: In the present in vitro study, all specimens were able to withstand 30° off-axis loading with the exception of one specimen. With these results, this would indicate that these denture teeth are able to withstand normal occlusal forces. [source]


Load-bearing capacity of all-ceramic posterior four-unit fixed partial dentures with different zirconia frameworks

EUROPEAN JOURNAL OF ORAL SCIENCES, Issue 2 2007
Philipp Kohorst
The aim of this in vitro study was to compare the load-bearing capacity of posterior four-unit fixed partial dentures (FPDs) produced with two different yttria-stabilized polycrystalline tetragonal zirconia (Y-TZP) ceramics, one being a presintered material, the other a fully sintered, hot isostatically pressed material. Additionally, as a novel approach, the influence of preliminary mechanical damage upon the fracture force of an FPD has been investigated. A total of 20 frameworks each were milled from presintered zirconia and from fully sintered zirconia. Prior to veneering, 10 frameworks of each material were ,damaged' by a defined saw cut similar to an accidental flaw generated during shape cutting. Before fracture testing, all FPDs were subjected to thermal and mechanical cycling. Additionally, scanning electron microscopy was used to investigate fracture surfaces. Statistical analysis showed that FPDs milled from fully sintered zirconia had a significantly higher fracture resistance compared with specimens made from presintered material, whereas preliminary damage did not have a significant effect. After aging, FPDs made from both materials were capable of withstanding occlusal forces reported in the literature. Therefore, both types of Y-TZP may be suitable for posterior four-unit all-ceramic FPDs, although further prolonged aging experiments and prospective clinical trials are required to prove their fitness for clinical use. [source]


Influence of food consistency on the rabbit masseter muscle fibres

EUROPEAN JOURNAL OF ORAL SCIENCES, Issue 1 2003
Geerling Langenbach
The plasticity of the masseter muscle was studied by comparing two groups of rabbits that were fed soft- and hard-diet for 87 d. Incisors of the soft-diet group were cut back to minimize the bite forces. Muscle fibres were immunohistochemically defined as fast- or slow-contracting fibres and their cross-sectional area was measured. The muscles of animals fed with the hard-diet were composed of fibres with larger cross-sectional areas than the soft-diet group. The relative difference was larger in slow-contracting fibres than in fast-contracting fibres. The results were similar for the different regions of the muscle. No changes in fibre composition were found. In conclusion, the difference in food consistency, as induced in this study, caused changes in the muscle fibre cross-sectional area that can be recognized from the altered necessary occlusal forces, which result from the modified forces developed by the masseter muscle. [source]


Ceramometal Bonded Inlays and Onlays

JOURNAL OF ESTHETIC AND RESTORATIVE DENTISTRY, Issue 3 2000
HARRY F. ALBERS DDS
ABSTRACT Porcelain-fused-to-metal (PFM) is presently the most popular crown and bridge technique. Its success is attributable to the use of a metal substructure to support the porcelain, resulting in good esthetics, limited ceramic flexure, and better longevity than all-porcelain restorations. A new concept suggests encapsulating the metal substructure with ceramic and placing the insert under the occlusal surface, thereby combining the best properties of the esthetics of porcelain and the support of metal to achieve improved restorations. CLINICAL SIGNIFICANCE The ceramometal bonded restoration (CMBR) addresses the known limitations of PFMs by providing a more esthetic, more conservative, and more durable alternative. The CMBR eliminates the potential for tooth darkening, leaves more of the natural tooth intact, and offers good resistance to fracture or displacement under occlusal forces. This combination of features makes the CMBR inlay or onlay an excellent alternative for the conservative esthetic restoration of posterior teeth. [source]


Tooth displacement due to occlusal contacts: a three-dimensional finite element study

JOURNAL OF ORAL REHABILITATION, Issue 12 2006
S. GOMES DE OLIVEIRA
summary, The use of the Finite Element Method (FE) is an appropriate way to study occlusal forces and tooth movement. The purpose of this study was to evaluate the effects of different occlusal contact patterns on tooth displacement in an adult dentition using a three-dimensional FE model of a human maxilla and mandible. Initially, images of a computerized tomography scan were redrawn in a computer program (CATIA) followed by the FE mesh construction. The MSC/Patran software was used to develop the FE mesh comprising 520 445 elements and 106 633 nodes. The MSC/Nastran program was utilized as pre and post-processor for all mathematical calculations necessary to evaluate dental and mandibular biomechanics. Four occlusal patterns were tested: FEM 1 , standard occlusal contacts; FEM 2 , removal of mesial marginal and mesial tripoidism contacts; FEM 3 , removal of distal marginal and distal tripoidism contacts; FEM 4 , similar to FEM 3 with added contacts between upper and lower incisors. Small changes in the standard distribution of occlusal contacts resulted in an imbalance of occlusal forces and changes in dental positioning. All simulations tested showed mesial displacement of posterior teeth. The most significant changes were registered in the model presenting unstable occlusal contacts when the anterior teeth were in occlusion (FEM 4). These findings may explain mandibular incisors crowding and maxillary incisors flaring as a result of small variations in dental contacts. [source]


A biomechanical effect of wide implant placement and offset placement of three implants in the posterior partially edentulous region

JOURNAL OF ORAL REHABILITATION, Issue 1 2000
Y. Sato
To prevent loosening or fracture of screws retaining the prosthesis to the implants in the posterior partially edentulous region, the use of staggered buccal and lingual offset placement or wide implants is suggested. However, it is not known how this usage compensates for the torque produced by lateral occlusal forces. This study evaluated the effectiveness of offset placement of three implants and a wide implant placement at the most posterior site. Three-dimensional geometric analysis was used to calculate the tensile force applied to gold screws in clinical situations with buccal or lingual loading perpendicular to cuspal inclination (10 or 20°). Four variations of the placement of three implants (, 3·75 mm) are: (1) straight; (2) buccal offset of the second implant; (3) lingual offset of the second implant; (4) a wide implant (, 5 mm) placement at most posterior site. The offset placement did not always decrease tensile force at the gold screw, but wide implant placement and decrease in cuspal inclination did. [source]


Failure Modes with Point Loading of Three Commercially Available Denture Teeth

JOURNAL OF PROSTHODONTICS, Issue 6 2008
Andrew R. Moffitt DDS
Abstract Purpose: A common problem associated with implant-supported prostheses is the fracture of denture teeth. This study was designed to compare the fracture modes of three denture teeth by compressive load at a 30° off-axis angle. Material and Methods: Three denture teeth (Vident Duostat, Ivoclar Vivadent, and Dentsply Trubyte) processed to two denture base processing systems [injection-molded (IM) SR-Ivocap system and compression-molded (CM) denture base resin] were evaluated. Each specimen was processed to a metal framework. Ultimate failure strength of each system when point loaded at a 30° off-axis angle was recorded, along with a visual inspection of each specimen. Results: The average load fracture for each group was (in N): Vident CM 1106.97 ± 223.20, Vident IM 1168.18 ± 322.52, Dentsply CM 1098.08 ± 286.32, Dentsply IM 1023.80 ± 282.45, Ivoclar CM 1616.98 ± 204.87, and Ivoclar IM 1373.54 ± 282.58. There was a significant difference between the groups and the Ivoclar CM group. The Ivoclar CM group had the highest average load force, and the Dentsply IM group had the lowest average load force. On average, the teeth within the groups fractured at a higher compression force than the average maximum occlusal force in natural dentition. Dentsply and Vident denture teeth fractured more horizontally, and the Ivoclar denture teeth fractured more vertically within the groups. There was no significant difference among the groups between the IM and CM processing methods. Conclusions: In the present in vitro study, all specimens were able to withstand 30° off-axis loading with the exception of one specimen. With these results, this would indicate that these denture teeth are able to withstand normal occlusal forces. [source]


Influence of implant diameter on surrounding bone

CLINICAL ORAL IMPLANTS RESEARCH, Issue 5 2007
Jeff Brink
Abstract Objectives: Implant osseointegration is dependent upon various factors, such as bone quality and type of implant surface. It is also subject to adaptation in response to changes in bone metabolism or transmission of masticatory forces. Understanding of long-term physiologic adjustment is critical to prevention of potential loss of osseointegration, especially because excessive occlusal forces lead to failure. To address this issue, wide-diameter implants were introduced in part with the hope that greater total implant surface would offer mechanical resistance. Yet, there is little evidence that variation in diameter translates into a different bone response in the implant vicinity. Therefore, this study aimed at comparing the impact of implant diameter on surrounding bone. Material and methods: Twenty standard (3.75 mm) and 20 wide (5 mm) implants were placed using an animal model. Histomorphometry was performed to establish initial bone density (IBD), bone to implant contact (BIC) and adjacent bone density (ABD). Results: BIC was 71% and 73%, whereas ABD was 65% and 52%, for standard and wide implants, respectively. These differences were not statistically different (P>0.05). Correlation with IBD was then investigated. BIC was not correlated with IBD. ABD was not correlated to IBD for standard implants (r2=0.126), but it was correlated with wide implants (r2=0.82). In addition, a 1 : 1 ratio between IBD and ABD was found for wide implants. It can be concluded, within the limits of this study, that ABD may be influenced by implant diameter, perhaps due to differences in force dissipation. [source]


Comparison between jaw bone augmentation by means of a stiff occlusive titanium membrane or an autologous hip graft: a retrospective clinical assessment

CLINICAL ORAL IMPLANTS RESEARCH, Issue 5 2006
Liene Molly
Abstract Objectives: Patients in need of bone augmentation procedures can be treated with a traditional hip graft or by neogenesis under a stiff occlusive titanium membrane, custom-fit on the basis of CT-scan data. Material and methods: Eighteen patients treated by means of a hip graft (H) were compared with 11 patients using a full titanium membrane (M) for the rehabilitation of upper jaws. In the H group, 17 patients (62 implants) in the anterior region (Ha) and 14 patients (23 implants) in the posterior region (Hp) were included. In the M group, nine patients (30 implants) in the anterior region (Ma) and seven patients (16 implants) in the posterior region (Mp) were included. Results: In group Ha, the CFR was 13.3% after 14 years compared with group Ma where the CFR was 17.4% after 9 years. In group Hp, the CFR was 22.8% after 16 years compared with group Mp where the CFR was 23.4% after 6 years. The marginal bone loss for group Ha after 20 years was 2.7 mm, for group Ma it was not even 1 mm after 9 years; for group Hp, it was 2.5 mm after 15 years compared with less than 1 mm after 3 years in group Mp. Conclusions: The success rate of alveolar ridge augmentation therapy is lower than in a classical approach. These augmentation procedures, however, show acceptable results. The titanium membrane augmentation leads to less marginal bone loss probably because neo-formed bone has a better chance to adapt its mineralization to occlusal forces encountered. This technique could offer improvement if membrane exposure can be avoided. [source]


Human ex vivo bone tissue strains around immediately loaded implants supporting maxillary overdentures

CLINICAL ORAL IMPLANTS RESEARCH, Issue 6 2005
vanç Akça
Abstract Purpose: To evaluate ex vivo bone tissue strains around maxillary implants supporting immediately loaded bar-retained overdentures. Material and methods: Ninety degree two-element rosette strain gauges were bonded on the labial bone of four ITI® dental implants placed in the maxillary anterior region of four completely edentulous maxilla of fresh human cadavers. The installation torque value (ITV) of each implant was measured by a custom-made torque wrench and resonance frequency analyses (RFAs) were undertaken. A bar-retained overdenture was fabricated for each cadaver, and two miniature load cells were integrated in the first molar region of the overdentures for controlled loading experiments. Strain measurements were performed at a sample rate of 10 kHz and under a maximum load of 100 N, simultaneously monitored from a computer connected to a data acquisition system. Finally, removal torque values (RTVs) of the implants were measured. Results: RFA values did not mirror ITVs, while RTVs of implants were slightly lower than the ITVs. Any correlation could not be obtained between RFA values and ITVs or RTVs. Maximum strains around loaded implants ranged between ,100 and ,550 ,, under 25,100 N. The axial and lateral strain values of posterior implants of both sides were higher than those of anterior implants under all loads (P<0.05). Conclusion: Because occlusal forces in humans tend to decrease because of age-related factors, maximum strains around immediately loaded implants supporting maxillary overdentures fall within physiologic levels. Résumé Le but de cette étude a été d'évaluer les tensions du tissu osseux ex vivo autour d'implants maxillaires portant des prothèses amovibles retenues sur une barre placée immédiatement après l'insertion des implants. Deux jauges de force en rosette à 90° ont été attachées au côté lingual de l'os de quatre implants dentaires ITI® placés dans la région antérieure du maxillaire de quatre cadavres humains frais aux maxillaires édentées. La valeur du couple de torsion lors de l'installation (ITV) de chaque implant a été mesurée par un couple fabriqué et des analyses de fréquence de résonnance (RFA) ont été faites. Une prothèse retenue par une barre a été fabriquée pour chaque cadavre et deux cellules de charge miniatures ont été intégrées dans la région de la première molaire de ces prothèses pour les expériences de charge contrôlées. Les mesures de force ont été effectuées à un taux d'échantillonnage de 10 kHz et sous une charge maximale de 100 N, suivies simultanément par un ordinateur connectéà un système d'acquisition de données. Finalement les valeurs des couples de torsion à l'enlèvement (RTV) des implants ont été mesurées. Les valeurs RFA n'étaient pas en ligne avec les ITV tandis que les RTV des implants étaient légèrement inférieurs aux ITV. Aucune corrélation n'a pûêtre établie entre les valeurs RFA et ITV ou RTV. Les forces maximales autour des implants chargés s'étalaient de ,100 ,e à,550 ,e sous 25 N à 100 N. Les valeurs des forces axiales et latérales des implants postérieurs des deux côtés étaient supérieures à celles des implants antérieurs sous toutes les charges (P<0.05). Zusammenfassung Ziel: Es war das Ziel dieser Arbeit, an bis vor kurzem vitalen Knochengewebe Spannungen um Oberkieferimplantate zu untersuchen, welche sofortbelastete und stegverankerte Hybridprothesen tragen. Material und Methoden: Auf den labialen Knochen von vier ITI-Implantaten in der vorderen Region des zahnlosen Oberkiefers von vier frisch verstorbenen Menschen klebte man Dehnmessinstrumente an. Man mass den Wert der Eindrehkraft (ITV) jedes einzelnen Implantates mit einem handelsblichen Drehmomentschlssel und fhrte eine Resonanzfrequenzanalyse (RFA) durch. Man stellte fr jede Leiche eine stegverankerte Hybridprothese her und man baute in der Region der ersten Molaren zwei kleine Messgerte ein, die kontrollierte Belastungsexperimente erlaubten. Dehnmessungen fhrte man mit einer Frequenz von 10 kHz und einer maximalen Belastung von 100 N durch, simultan aufgezeichnet von einem mit dem Messsystem verbundenen Komputer. Schliesslich mass man auch die Krfte, die es brauchte um die Implantate wieder auszudrehen (RTV). Resultate: Die RFA wiederspiegelte die ITV-Werte nicht, whrenddem die RTV-Werte der Implantate geringfgig tiefer waren als die ITV-Werte. Man konnte weder zwischen der RFA, den ITV-Werten oder den RTV-Werten eine Korrelation herstellen. Die maximalen Dehnungen um die Implantate erreichte Werte zwischen ,100 E und ,550 E bei einer Belastung zwischen 25 N und 100 N. Die axialen und lateralen Dehnkrfte bei posterioren Implantaten waren unter smtlichen getesteten Belastungen beidseits grsser als diejenigen der anterioren Implantate (P<0.05). Resumen Propósito: Evaluar las tensiones del tejido óseo ex vivo alrededor de implantes maxilares soportando sobredentaduras retenidas por barras con carga inmediata. Material y Métodos: Se pegaron dos indicadores de tensión de roseta de dos elementos de 90° en el hueso labial de cuatro implantes dentales ITI® colocados en la región maxilar anterior de cuatro maxilares completamente edéntulos de cadáveres humanos frescos. Se midió el valor del torque de instalación (ITV) de cada implante por medio de una chicharra de torque hecha a medida y se llevó a cabo análisis de frecuencia de resonancia (RFA). Se fabricó una sobredentadura retenida por barras para cada cadáver y se integraron dos células miniatura de carga en la región del primer molar de las sobredentaduras para los experimentos de carga controlada. Se llevaron a cabo mediciones de la tensión a un índice de muestra de 10 kHz y bajo una carga máxima de 100 N, simultáneamente monitorizada desde un ordenador conectado a sistema de adquisición de datos. Finalmente, se midieron los valores del torque de remoción (RTV) de los implantes. Resultados: Los valores de RFA no se reflejaron en los ITVs, mientras que los RTVs de los implantes fueron ligeramente más bajos que los ITVs. No se pudieron obtener correlaciones entre los valores de RFA y los ITVs o RTVE. Las tensiones máximas alrededor de los implantes cargados variaron entre ,100 ,, y ,550 ,, bajo 25 N a 100 N. Los valores de las tensiones axiales y laterales de los implantes posteriores de ambos lados fueron mayores que aquellos implantes anteriores bajo todas las cargas (P<0.05). [source]