High Mass Accuracy (high + mass_accuracy)

Distribution by Scientific Domains


Selected Abstracts


Glycoform characterization of erythropoietin combining glycan and intact protein analysis by capillary electrophoresis , electrospray , time-of-flight mass spectrometry

ELECTROPHORESIS, Issue 13 2006
Elvira Balaguer
Abstract Glycosylation of recombinant human erythropoietin (rHuEPO) is a post-translational process that alters biological activity, solubility and lifetime of the glycoprotein in blood, and strongly depends on the type of cell and the cell culture conditions. A fast and simple method providing extensive carbohydrate information about the glycans present in rHuEPO and other glycoproteins is needed in order to improve current methods in drug development or product quality control. Here, an improved method for intact rHuEPO glycoform characterization by CZE-ESI-TOF MS has been developed using a novel capillary coating and compared to a previous study. Both methods allow a fast separation in combination with accurate mass characterization of the single protein isoforms. The novel dynamic coating provides a separation at an EOF close to zero, enabling better separation. This results in an improved mass spectrometric resolution and the detection of minor isoforms. In order to assign an unequivocal carbohydrate composition to every intact glycoform, a CZE-ESI-MS separation method for enzymatically released underivatized N -glycans has been developed. The TOF,MS allows the correct identification of the glycans due to its high mass accuracy and resolution. Therefore, glycan modifications such as acetylation, oxidation, sulfation and even the exchange of OH by NH2 are successfully characterized. Information of the protein-backbone molecular mass has been combined with results from peptide analysis (revealing information about O -glycosylation) and from the glycan analysis, including the detection of as yet undescribed glycans containing four antennae and five sialic acids. This allows an unequivocal assignment of an overall glycosylation composition to the molecular masses obtained for the intact rHuEPO glycoforms. [source]


The Orbitrap: a new mass spectrometer

JOURNAL OF MASS SPECTROMETRY (INCORP BIOLOGICAL MASS SPECTROMETRY), Issue 4 2005
Qizhi Hu
Abstract Research areas such as proteomics and metabolomics are driving the demand for mass spectrometers that have high performance but modest power requirements, size, and cost. This paper describes such an instrument, the Orbitrap, based on a new type of mass analyzer invented by Makarov. The Orbitrap operates by radially trapping ions about a central spindle electrode. An outer barrel-like electrode is coaxial with the inner spindlelike electrode and mass/charge values are measured from the frequency of harmonic ion oscillations, along the axis of the electric field, undergone by the orbitally trapped ions. This axial frequency is independent of the energy and spatial spread of the ions. Ion frequencies are measured non-destructively by acquisition of time-domain image current transients, with subsequent fast Fourier transforms (FFTs) being used to obtain the mass spectra. In addition to describing the Orbitrap mass analyzer, this paper also describes a complete Orbitrap-based mass spectrometer, equipped with an electrospray ionization source (ESI). Ions are transferred from the ESI source through three stages of differential pumping using RF guide quadrupoles. The third quadrupole, pressurized to less than 10,3 Torr with collision gas, acts as an ion accumulator; ion/neutral collisions slow the ions and cause them to pool in an axial potential well at the end of the quadrupole. Ion bunches are injected from this pool into the Orbitrap analyzer for mass analysis. The ion injection process is described in a simplified way, including a description of electrodynamic squeezing, field compensation for the effects of the ion injection slit, and criteria for orbital stability. Features of the Orbitrap at its present stage of development include high mass resolution (up to 150 000), large space charge capacity, high mass accuracy (2,5 ppm), a mass/charge range of at least 6000, and dynamic range greater than 10.3 Applications based on electrospray ionization are described, including characterization of transition-metal complexes, oligosaccharides, peptides, and proteins. Use is also made of the high-resolution capabilities of the Orbitrap to confirm the presence of metaclusters of serine octamers in ESI mass spectra and to perform H/D exchange experiments on these ions in the storage quadrupole. Copyright © 2005 John Wiley & Sons, Ltd. [source]


The minotaur proteome: Avoiding cross-species identifications deriving from bovine serum in cell culture models

PROTEINS: STRUCTURE, FUNCTION AND BIOINFORMATICS, Issue 16 2010
Jakob Bunkenborg
Abstract Cell culture is a fundamental tool in proteomics where mammalian cells are cultured in vitro using a growth medium often supplemented with 5,15% FBS. Contamination by bovine proteins is difficult to avoid because of adherence to the plastic vessel and the cultured cells. We have generated peptides from bovine serum using four sample preparation methods and analyzed the peptides by high mass accuracy LC-MS/MS. Distinguishing between bovine and human peptides is difficult because of a considerable overlap of identical tryptic peptide sequences. Pitfalls in interpretation, different database search strategies to minimize erroneous identifications and an augmented contaminant database are presented. [source]


De novo sequencing of peptides by MS/MS

PROTEINS: STRUCTURE, FUNCTION AND BIOINFORMATICS, Issue 4 2010
Joerg Seidler
Abstract The current status of de novo sequencing of peptides by MS/MS is reviewed with focus on collision cell MS/MS spectra. The relation between peptide structure and observed fragment ion series is discussed and the exhaustive extraction of sequence information from CID spectra of protonated peptide ions is described. The partial redundancy of the extracted sequence information and a high mass accuracy are recognized as key parameters for dependable de novo sequencing by MS. In addition, the benefits of special techniques enhancing the generation of long uninterrupted fragment ion series for de novo peptide sequencing are highlighted. Among these are terminal 18O labeling, MSn of sodiated peptide ions, N-terminal derivatization, the use of special proteases, and time-delayed fragmentation. The emerging electron transfer dissociation technique and the recent progress of MALDI techniques for intact protein sequencing are covered. Finally, the integration of bioinformatic tools into peptide de novo sequencing is demonstrated. [source]


A strategy for high-resolution protein identification in surface-enhanced laser desorption/ionization mass spectrometry: Calgranulin A and chaperonin 10 as protein markers for endometrial carcinoma

PROTEINS: STRUCTURE, FUNCTION AND BIOINFORMATICS, Issue 7 2005
Jingzhong Guo
Abstract Surface-enhanced laser desorption/ionization-mass spectrometry (SELDI-MS) has conventionally been practiced on linear time of flight (TOF) which has low mass accuracy and resolution. Here we demonstrate in an examination of both malignant and nonmalignant endometrial tissue homogenates that high mass accuracy and resolution in the MS stage are crucial. Using a commercially available quadrupole/TOF (QqTOF), we were able to resolve two potential cancer markers, subsequently identified off-line as chaperonin 10 and calgranulin A, that differ by 8 Da in mass. Two off-line protein identification protocols were developed: the first was based on size-exclusion chromatography (SEC), sodium dodecyl sulfate-polyacrylamide gel electrophoresis (SDS-PAGE), protein extraction, trypsin digestion, and matrix-assisted laser desorption/ionization-tandem MS (MALDI-MS/MS); the second on SEC and shotgun nano-liquid chromatography (nanoLC)-MS/MS. Analyses on a cohort of 44 endometrial homogenates showed 22 out of 23 nonmalignant samples had nondetectable to very low abundance of chaperonin 10 and calgranulin A; 17 of the 21 malignant samples had detectable to abundant levels of both proteins. Immunohistochemical staining of a tissue microarray of 32 samples showed that approximately half of malignant endometrial tissues exhibited positive staining for calgranulin A in the malignant epithelium, while 9 out of 10 benign tissues exhibited negative epithelial staining. In addition, macrophages/granulocytes in malignant as well as nonmalignant tissues showed positive staining. No immunostaining occurred in stroma or myometrium. Calgranulin A, in combination with chaperonin 10 and other proteins, may eventually constitute a panel of markers to permit diagnosis and screening of endometrial cancer. [source]


High-resolution extracted ion chromatography, a new tool for metabolomics and lipidomics using a second-generation orbitrap mass spectrometer

RAPID COMMUNICATIONS IN MASS SPECTROMETRY, Issue 10 2009
Albert Koulman
Most analytical methods in metabolomics are based on one of two strategies. The first strategy is aimed at specifically analysing a limited number of known metabolites or compound classes. Alternatively, an unbiased approach can be used for profiling as many features as possible in a given metabolome without prior knowledge of the identity of these features. Using high-resolution mass spectrometry with instruments capable of measuring m/z ratios with sufficiently low mass measurement uncertainties and simultaneous high scan speeds, it is possible to combine these two strategies, allowing unbiased profiling of biological samples and targeted analysis of specific compounds at the same time without compromises. Such high mass accuracy and mass resolving power reduces the number of candidate metabolites occupying the same retention time and m/z ratio space to a minimum. In this study, we demonstrate how targeted analysis of phospholipids as well as unbiased profiling is achievable using a benchtop orbitrap instrument after high-speed reversed-phase chromatography. The ability to apply both strategies in one experiment is an important step forward in comprehensive analysis of the metabolome. Copyright © 2009 John Wiley & Sons, Ltd. [source]


,All-in-One' analysis for metabolite identification using liquid chromatography/hybrid quadrupole time-of-flight mass spectrometry with collision energy switching

RAPID COMMUNICATIONS IN MASS SPECTROMETRY, Issue 18 2005
Mark Wrona
The removal of bottlenecks in discovery stage metabolite identification studies is an ongoing challenge for the pharmaceutical industry. We describe the use of an ,All-in-One' approach to metabolite characterization that leverages the fast scanning and high mass accuracy of hybrid quadrupole time-of-flight mass spectrometry (QqToFMS) instruments. Full-scan MS and MS/MS data is acquired using collision energy switching without the preselection, either manually or in a data-dependent manner, of precursor ions. The acquisition of ,clean' MS/MS data is assisted by the use of ultrahigh-performance chromatography. Data acquired using this method can then be mined post-acquisition in a number of ways. These include using narrow window extracted ion chromatograms (nwXICs) for expected biotransformations, XICs for the product ions of the parent compound and/or expected modification of these product ions, and neutral loss chromatograms. This approach has the potential to be truly comprehensive for the determination of in vitro biotransformations in a drug discovery environment. Copyright © 2005 John Wiley & Sons, Ltd. [source]


Mass spectrometry for the detection of differentially expressed proteins: a comparison of surface-enhanced laser desorption/ionization and capillary electrophoresis/mass spectrometry

RAPID COMMUNICATIONS IN MASS SPECTROMETRY, Issue 2 2004
Nils v. Neuhoff
The discovery of biomarkers is currently attracting much interest as it harbors great potential for the diagnosis and monitoring of human diseases. Here we have used two advanced mass spectroscopy based technologies, surface enhanced laser desorption ionization (SELDI-MS) and capillary electrophoresis/mass spectrometry (CE/MS), to obtain proteomic patterns of urine samples from patients suffering from membranous glomerulonephritis (MGN) and healthy volunteers. The results indicate that CE/MS analysis is able to display a rich and complex pattern of polypeptides with high resolution and high mass accuracy. In order to analyze these patterns, the MosaiqueVisu software was developed for peak identification, deconvolution and the display of refined maps in a three-dimensional format. The polypeptide profiles obtained with SELDI-MS from the same samples are much sparser and show lower resolution and mass accuracy. The SELDI-MS profiles are further heavily dependent on analyte concentration. SELDI-MS analysis identified three differentially expressed polypeptides, which are potential biomarkers that can distinguish healthy donors from patients with MGN. In contrast, approximately 200 potential biomarkers could be identified by CE/MS. Thus, while SELDI-MS is easy to use and requires very little sample, CE/MS generates much richer data sets that enable an in-depth analysis. Copyright © 2003 John Wiley & Sons, Ltd. [source]


Closely spaced external standard: a universal method of achieving 5 ppm mass accuracy over the entire MALDI plate in axial matrix-assisted laser desorption/ionization time-of-flight mass spectrometry

RAPID COMMUNICATIONS IN MASS SPECTROMETRY, Issue 19 2003
Eugene Moskovets
Close deposition of the sample and external standard was used in axial matrix-assisted laser desorption/ionization time-of-flight mass spectrometry (MALDI-TOFMS) to achieve mass accuracy equivalent to that obtained with an internal standard across the entire MALDI plate. In this work, the sample and external standard were deposited by continuous deposition in separate traces, each approximately 200,,m wide. The dependence of the mass accuracy on the distance between the sample and standard traces was determined across a MALDI target plate with dimensions of 57.5,mm,×,57.0,mm by varying the gap between the traces from 100,,m to 4,mm. During acquisition, two adjacent traces were alternately irradiated with a 200-Hz laser, such that the peaks in the resulting mass spectra combined the sample and external standard. Ion suppression was not observed even when the peptide concentrations in the two traces differed by more than two orders of magnitude. The five peaks from the external standard trace were used in a four-term mass calibration of the masses of the sample trace. The average accuracy across the whole plate with this method was 5,ppm when peaks of the sample trace had signal-to-noise ratios of at least 30 and the gap between the traces was approximately 100,,m. This approach was applied to determining peptide masses of a reversed-phase liquid chromatographic (LC) separation of a tryptic digest of , -galactosidase deposited as a long serpentine trace across the MALDI plate, with accuracy comparable to that obtainable using internal calibration. In addition, the eluent from reversed-phase LC separation of a strong cation-exchange fraction containing tryptic peptides from a yeast lysate along with the closely placed external standard was deposited on the MALDI plate. The data obtained in the MS and MS/MS modes on a MALDI-TOF/TOF mass spectrometer were combined and used in database searching with MASCOT. Since the significant score is a function of mass accuracy in the MS mode, database searching with high mass accuracy reduced the number of false positives and also added peptides which otherwise would have been eliminated at lower mass accuracy (false negatives). Copyright © 2003 John Wiley & Sons, Ltd. [source]


Primary structural determination of N-terminally blocked peptides from the bark of Eucommia ulmoides Oliv by mass spectrometric analysis

RAPID COMMUNICATIONS IN MASS SPECTROMETRY, Issue 9 2003
Ren-Huai Huang
Sequencing of N-terminally blocked proteins/peptides is a challenge since these molecules inhibit processing by Edman degradation. By using high accuracy Fourier transform ion cyclotron resonance (FTICR) tandem mass spectrometry and matrix-assisted laser desorption/ionization time-of-flight mass spectrometry (MALDI-TOFMS), the primary structures of two novel N-terminally blocked antifungal peptides (EAFP1 and EAFP2), purified from the bark of Eucommia ulmoides Oliv, have been determined. The results show that the high mass accuracy provided by FTICR mass spectrometry is effective to determine the N-terminally blocking group, and can simplify the task of spectral interpretation and improve the precision of sequence determination. The combination of MALDI-TOFMS with carboxyl peptidase Y digestion was used to determine the C-terminal 36- and 27-residue sequences of EAFP1 and EAFP2, respectively, to provide the sequence linkage information for tryptic fragments. Compared with traditional peptide chemistry the advantage of mass spectrometric techniques is their simplicity, speed and sensitivity. Copyright © 2003 John Wiley & Sons, Ltd. [source]