Surface ECG (surface + ecg)

Distribution by Scientific Domains


Selected Abstracts


Different Narrow QRS Morphologies in the Surface ECG: What is the Mechanism?

JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, Issue 8 2009
LUCIO CAPULZINI M.D.
No abstract is available for this article. [source]


Use of an Intracardiac Electrogram Eliminates the Need for a Surface ECG during Implantable Cardioverter-Defibrillator Follow-Up

PACING AND CLINICAL ELECTROPHYSIOLOGY, Issue 12 2007
KEVIN A. MICHAEL M.B.Ch.B.
Background:A surface electrocardiogram (SECG) for pacing threshold measurements during routine implantable cardioverter-defibrillator (ICD) follow-up can be cumbersome. This study evaluated the use of an intrathoracic far-field electrogram (EGM) derived between the Can and superior vena cava (SVC) electrode,the Leadless electrocardiogram (LLECG), in dual chamber ICDs in performing pacing threshold tests. Methods:The LLECG was evaluated prospectively during atrial and ventricular pacing threshold testing as a substudy of the Comparison of Empiric to Physician-Tailored Programming of Implantable Cardioverter-Defibrillators trial (EMPIRIC) in which dual chamber ICDs were implanted in 888 patients. Threshold tests were conducted at 1 volt by decrementing the pulse width. Follow-up at three months compared pacing thresholds measured using LLECG with those using Lead I of the surface ECG (SECG). The timesaving afforded by LLECG was assessed by a questionnaire. Results:The median threshold difference between LLECG and SECG measurements for both atrial (0.00 ms, P = 0.90) and ventricular (0.00 ms, P = 0.34) threshold tests were not significant. Ninety percent of atrial and ventricular threshold differences were bounded by ± 0.10 ms and ,0.10 to +0.04 ms, respectively. We found that 99% of atrial and ventricular thresholds tests at six and 12 months attempted using LLECG were successfully completed. The questionnaire indicated that 65% of healthcare professionals found LLECG to afford at least some timesaving during device follow-ups. Conclusion:Routine follow-up can be performed reliably and expeditiously in dual chamber Medtronic (Minneapolis, MN, USA) ICDs using LLECG alone, resulting in overall timesaving. [source]


R-Wave Detection by Subcutaneous ECG.

ANNALS OF NONINVASIVE ELECTROCARDIOLOGY, Issue 1 2001
Possible Use for Analyzing R-R Variability
Background: Atrial arrhythmia (AA) discrimination remains a technological challenge for implanted cardiac devices. We examined the feasibility of R-wave detection by a subcutaneous far field ECG (SFFECG) and analysis of these signals for R to R variability as an indicator of atrial arrhythmia (AA). Methods: Surface ECG and SFFECG (from the pacemaker pocket) were recorded in sixteen patients (61.5 ± 11.4 years) with AA. The SFFECG was recorded with a pacemaker sized four electrode array acutely placed in the pacemaker pocket during implantation. The signals were analyzed to obtain peak-to-peak R wave amplitude and R to R interval variability (indicative of AAs). Results: In sixteen patients R waves were visually discernible in all recordings. The percentage over and under detection for automatic R wave recognition SFFECG was 3 and 9%, respectively. R to R variability analysis using the SFFECG produced results concordant to those using the surface ECG. Conclusion: SFFECG might be a helpful adjunct in implantable device systems for detection of R waves and may be used for measurement of R to R variability. A.N.E. 2001;6(1):18,23 [source]


Effects of an adapted intravenous amiodarone treatment protocol in horses with atrial fibrillation

EQUINE VETERINARY JOURNAL, Issue 4 2007
D. de CLERCQ
Summary Reason for performing study: Good results have been obtained with a human amiodarone (AD) i.v. protocol in horses with chronic atrial fibrillation (AF) and a pharmacokinetic study is required for a specific i.v. amiodarone treatment protocol for horses. Objectives: To study the efficacy of this pharmacokinetic based i.v. AD protocol in horses with chronic AF. Methods: Six horses with chronic AF were treated with an adapted AD infusion protocol. The protocol consisted of 2 phases with a loading dose followed by a maintenance infusion. In the first phase, horses received an infusion of 6.52 mg AD/kg bwt/h for 1 h followed by 1.1 mg/kg bwt/h for 47 h. In the second phase, horses received a second loading dose of 3.74 mg AD/kg bwt/h for 1 h followed by 1.31 mg/kg bwt/h for 47 h. Clinical signs were monitored, a surface ECG and an intra-atrial electrogram were recorded. AD treatment was discontinued when conversion or any side effects were observed. Results: Three of the 6 horses cardioverted successfully without side effects. The other 3 horses did not convert and showed adverse effects, including diarrhoea. In the latter, there were no important circulatory problems, but the diarrhoea continued for 10,14 days. The third horse had to be subjected to euthanasia because a concomitant Salmonella infection worsened the clinical signs. Conclusion: The applied treatment protocol based upon pharmacokinetic data achieved clinically relevant concentrations of AD and desethylamiodarone. Potential relevance: Intravenous AD has the potential to be an alternative pharmacological treatment for AF in horses, although AD may lead to adverse drug effects, particularly with cumulative dosing. [source]


Preoperative Electrocardiographic Risk Assessment of Atrial Fibrillation After Coronary Artery Bypass Grafting

JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, Issue 12 2004
Ph.D., YI GANG M.D.
Introduction: This study evaluated the role of surface ECG in assessment of risk of new-onset atrial fibrillation (AF) after coronary artery bypass grafting surgery (CABG). Methods and Results: One hundred fifty-one patients (126 men and 25 women; age 65 ± 10 years) without a history of AF undergoing primary elective and isolated CABG were studied. Standard 12-lead ECGs and P wave signal-averaged ECG (PSAE) were recorded 24 hours before CABG using a MAC VU ECG recorder. In addition to routine ECG measurements, two P wave (P wave complexity ratio [pCR]; P wave morphology dispersion [PMD]) and six T wave morphology descriptors (total cosine R to T [TCRT]; T wave morphology dispersion of ascending and descending part of the T wave [aTMD and dTMD], and others), and three PSAE indices (filtered P wave duration [PD]; root mean square voltage of terminal 20 msec of averaged P wave [RMS20]; and integral of P wave [Pi]) were investigated. During a mean hospital stay of 7.3 ± 6.2 days after CABG, 40 (26%) patients developed AF (AF group) and 111 remained AF-free (no AF group). AF patients were older (69 ± 9 years vs 64 ± 10 years, P = 0.005). PD (135 ± 9 msec vs 133 ± 12 msec, P = NS) and RMS20 (4.5 ± 1.7 ,V vs 4.0 ± 1.6 ,V, P = NS) in AF were similar to that in no AF, whereas Pi was significantly increased in AF (757 ± 230 ,Vmsec vs 659 ± 206 ,Vmsec, P = 0.007). Both pCR (32 ± 11 vs 27 ± 10) and PMD (31.5 ± 14.0 vs 26.4 ± 12.3) were significantly greater in AF (P = 0.012 and 0.048, respectively). TCRT (0.028 ± 0.596 vs 0.310 ± 0.542, P = 0.009) and dTMD (0.63 ± 0.03 vs 0.64 ± 0.02, P = 0.004) were significantly reduced in AF compared with no AF. Measurements of aTMD and three other T wave descriptors were similar in AF and no AF. Significant variables by univariate analysis, including advanced age (P = 0.014), impaired left ventricular function (P = 0.02), greater Pi (P = 0.012), and lower TCRT (P = 0.007) or dTMD, were entered into multiple logistic regression models. Increased Pi (P = 0.038), reduced TCRT (P = 0.040), and lower dTMD (P = 0.014) predicted AF after CABG independently. In patients <70 years, a linear combination of increased pCR and lower TCRT separated AF and no AF with a sensitivity of 74% and specificity of 62% (P = 0.005). Conclusion: ECG assessment identifies patients vulnerable to AF after CABG. Combination of ECG parameters assessed preoperatively may play an important role in predicting new-onset AF after CABG. [source]


Surface Atrial Frequency Analysis in Patients with Atrial Fibrillation:

JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, Issue 9 2004
A Tool For Evaluating the Effects of Intervention
Introduction: The aims of this study were to evaluate (1) principal component analysis as a technique for extracting the atrial signal waveform from the standard 12-lead ECG and (2) its ability to distinguish changes in atrial fibrillation (AF) frequency parameters over time and in response to pharmacologic manipulation using drugs with different effects on atrial electrophysiology. Methods and Results: Twenty patients with persistent AF were studied. Continuous 12-lead Holter ECGs were recorded for 60 minutes, first, in the drug-free state. Mean and variability of atrial waveform frequency were measured using an automated computer technique. This extracted the atrial signal by principal component analysis and identified the main frequency component using Fourier analysis. Patients were then allotted sequentially to receive 1 of 4 drugs intravenously (amiodarone, flecainide, sotalol, or metoprolol), and changes induced in mean and variability of atrial waveform frequency measured. Mean and variability of atrial waveform frequency did not differ within patients between the two 30-minute sections of the drug-free state. As hypothesized, significant changes in mean and variability of atrial waveform frequency were detected after manipulation with amiodarone (mean: 5.77 vs 4.86 Hz; variability: 0.55 vs 0.31 Hz), flecainide (mean: 5.33 vs 4.72 Hz; variability: 0.71 vs 0.31 Hz), and sotalol (mean: 5.94 vs 4.90 Hz; variability: 0.73 vs 0.40 Hz) but not with metoprolol (mean: 5.41 vs 5.17 Hz; variability: 0.81 vs 0.82 Hz). Conclusion: A technique for continuously analyzing atrial frequency characteristics of AF from the surface ECG has been developed and validated. [source]


Drug-Induced Torsades de Pointes and Implications for Drug Development

JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, Issue 4 2004
Ph.D., ROBERT R. FENICHEL M.D.
Torsades de pointes is a potentially lethal arrhythmia that occasionally appears as an adverse effect of pharmacotherapy. Recently developed understanding of the underlying electrophysiology allows better estimation of the drug-induced risks and explains the failures of older approaches through the surface ECG. This article expresses a consensus reached by an independent academic task force on the physiologic understanding of drug-induced repolarization changes, their preclinical and clinical evaluation, and the risk-to-benefit interpretation of drug-induced torsades de pointes. The consensus of the task force includes suggestions on how to evaluate the risk of torsades within drug development programs. Individual sections of the text discuss the techniques and limitations of methods directed at drug-related ion channel phenomena, investigations aimed at action potentials changes, preclinical studies of phenomena seen only in the whole (or nearly whole) heart, and interpretation of human ECGs obtained in clinical studies. The final section of the text discusses drug-induced torsades within the larger evaluation of drug-related risks and benefits. (J Cardiovasc Electrophysiol, Vol. 15, pp. 475-495, April 2004) [source]


Demonstration of Electrical and Anatomic Connections Between Marshall Bundles and Left Atrium in Dogs: Implications on the Generation of P Waves on Surface Electrocardiogram

JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, Issue 12 2002
CHIKAYA OMICHI M.D.
Marshall Bundle and P Wave.Introduction: The muscle bundles within the ligament of Marshall (LOM) are electrically active. The importance of these muscle bundles (Marshall bundle [MB]) to atrial activation and the generation of the ECG P wave is unclear. Methods and Results: We used optical mapping techniques to study epicardial activation patterns in isolated perfused left atrium in four dogs. In another seven dogs, P waves were studied before and after in vivo radiofrequency (RF) ablation of the connection between coronary sinus (CS) and the LOM. Computerized mapping was performed before and after RF ablation. Optical mapping studies showed that CS pacing resulted in broad wavefronts propagating from the middle and distal LOM directly to the adjacent left atrium (LA). Serial sections showed direct connection between MB and LA near the orifice of the left superior pulmonary vein in two dogs. In vivo studies showed that MB potentials were recorded in three dogs. After ablation, the duration of P waves remained unchanged. In the other four dogs, MB potentials were not recorded. Computerized mapping showed that LA wavefronts propagated to the MB region via LA-MB connection and then excited the CS. After ablation, the activation of CS muscle sleeves is delayed, and P wave duration increased from 65.3 ± 14.9 msec to 70.5 ± 17.2 msec (P = 0.025). Conclusion: In about half of the normal dogs, MB provides an electrical conduit between LA free wall and CS. Severing MB alters the atrial activation and lengthens the P wave. MB contributes to generation of the P wave on surface ECG. [source]


Arrhythmogenesis of T Wave Alternans Associated with Surface QRS Complex Alternans and the Role of Ventricular Prematurity: Observations from a Canine Model of LQT3 Syndrome

JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, Issue 6 2002
MASAOMI CHINUSHI M.D.
Intramural TWA and Its Arrhythmogenesis.Introduction: T wave alternans (TWA) is characterized by cycle-to-cycle changes in the QT interval and/or T wave morphology. It is believed to amplify the underlying dispersion of ventricular repolarization. The aim of this study was to examine the mechanisms and arrhythmogenesis of TWA accompanied by QRS complex and/or blood pressure (BP) waveform alternans, using transmural ventricular electrogram recordings in an anthopleurin-A model of long QT syndrome. Methods and Results: The cardiac cycle length was gradually shortened by interruption of vagal stimulation, and TWA was induced in six canine hearts. Transmural unipolar electrograms were recorded with plunge needle electrodes from endocardial (Endo), mid-myocardial (Mid), and epicardial (Epi) sites, along with the surface ECG and BP. The activation-recovery interval (ARI) was measured to estimate local refractoriness. During TWA, ARI alternans was greater at the Mid than the Epi/Endo sites, and it was associated with the development of marked spatial dispersion of ventricular repolarization. As TWA increased, ventricular activation of the cycles associated with shorter QT intervals displayed delayed conduction at the Mid sites as a result of a critically longer ARI of the preceding cycle and longer QT interval, while normal conduction was preserved at the Epi site. Delayed conduction at the Mid sites manifested as surface ECG QRS and BP waveform alternans, and spontaneous ventricular tachyarrhythmias developed in absence of ventricular prematurity. In other instances, in absence of delayed conduction during TWA, ventricular premature complexes infringed on a prominent spatial dispersion of ventricular repolarization of cycles with long QT intervals and initiated ventricular tachyarrhythmia. Conclusion: TWA accompanied by QRS alternans may signal a greater ventricular electrical instability, since it is associated with intramural delayed conduction, which can initiate ventricular tachyarrhythmia without ventricular premature complexes. [source]


Maturational Atrioventricular Nodal Physiology in the Mouse

JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, Issue 5 2000
COLIN T. MAGUIRE B.S.
Mouse AV Nodal Maturation. Introduction: Dual AV nodal physiology is characterized by discontinuous conduction from the atrium to His bundle during programmed atrial extrastimulus testing (A2V2 conduction curves), AV nodal echo beats, and induction of AV nodal reentry tachycardia (AVNRT). The purpose of this study was to characterize in vivo murine maturational AV nodal conduction properties and determine the frequency of dual AV nodal physiology and inducible AVNRT. Methods and Results: A complete transvenous in vivo electrophysiologic study was performed on 30 immature and 19 mature mice. Assessment of AV nodal conduction included (1) surface ECG and intracardiac atrial and ventricular electrograms; (2) decremental atrial pacing to the point of Wenckebach block and 2:1 conduction; and (3) programmed premature atrial extrastimuli to determine AV effective refractory periods (AVERP), construct A2V2 conduction curves, and attempt arrhythmia induction. The mean Wenckebach block interval was 73 ± 12 msec, 2:1 block pacing cycle length was 61 ± 11 msec, and mean AVERP100 was 54 ± 11 msec. The frequency of dual AV nodal physiology increased with chronologic age, with discontinuous A2V2, conduction curves or AV nodal echo heats in 27% of young mice < 8 weeks and 58% in adult mice (P = 0.03). Conclusion: These data suggest that mice, similar to humans, have maturation of AV nodal physiology, hut they do not have inducible AVNRT. Characterization of murine electrophysiology may be of value in studying genetically modified animals with AV conduction abnormalities. Furthermore, extrapolation to humans may help explain the relative rarity of AVNRT in the younger pediatric population. [source]


Use of an Intracardiac Electrogram Eliminates the Need for a Surface ECG during Implantable Cardioverter-Defibrillator Follow-Up

PACING AND CLINICAL ELECTROPHYSIOLOGY, Issue 12 2007
KEVIN A. MICHAEL M.B.Ch.B.
Background:A surface electrocardiogram (SECG) for pacing threshold measurements during routine implantable cardioverter-defibrillator (ICD) follow-up can be cumbersome. This study evaluated the use of an intrathoracic far-field electrogram (EGM) derived between the Can and superior vena cava (SVC) electrode,the Leadless electrocardiogram (LLECG), in dual chamber ICDs in performing pacing threshold tests. Methods:The LLECG was evaluated prospectively during atrial and ventricular pacing threshold testing as a substudy of the Comparison of Empiric to Physician-Tailored Programming of Implantable Cardioverter-Defibrillators trial (EMPIRIC) in which dual chamber ICDs were implanted in 888 patients. Threshold tests were conducted at 1 volt by decrementing the pulse width. Follow-up at three months compared pacing thresholds measured using LLECG with those using Lead I of the surface ECG (SECG). The timesaving afforded by LLECG was assessed by a questionnaire. Results:The median threshold difference between LLECG and SECG measurements for both atrial (0.00 ms, P = 0.90) and ventricular (0.00 ms, P = 0.34) threshold tests were not significant. Ninety percent of atrial and ventricular threshold differences were bounded by ± 0.10 ms and ,0.10 to +0.04 ms, respectively. We found that 99% of atrial and ventricular thresholds tests at six and 12 months attempted using LLECG were successfully completed. The questionnaire indicated that 65% of healthcare professionals found LLECG to afford at least some timesaving during device follow-ups. Conclusion:Routine follow-up can be performed reliably and expeditiously in dual chamber Medtronic (Minneapolis, MN, USA) ICDs using LLECG alone, resulting in overall timesaving. [source]


Epicardial Ablation of Ventricular Tachycardia Associated with Isolated Ventricular Noncompaction

PACING AND CLINICAL ELECTROPHYSIOLOGY, Issue 7 2006
HONG EUY LIM
A 52-year-old man presented with sudden onset of palpitations and dizziness. Echocardiogram confirmed the diagnosis of isolated noncompaction of ventricular myocardium with moderated systolic dysfunction, and the electrocardiogram (ECG) revealed ventricular tachycardia (VT), of which the focus seemed to match an area of prominent left ventricular noncompaction on the 12-lead surface ECG. Through the activation mapping from the endo- and epicardium, simultaneously, a discrete potential preceding the QRS during VT was observed at the anterolateral epicardial wall. He subsequently underwent radiofrequency ablation, and VT was successfully eliminated. [source]


Spatial Distribution of Repolarization Times in Patients with Coronary Artery Disease

PACING AND CLINICAL ELECTROPHYSIOLOGY, Issue 8 2003
PETER VAN LEEUWEN
The potential clinical value of QT dispersion (QTd), a measure of the interlead range of QT interval duration in the surface 12-lead ECG, remains ambiguous. The aim of the study was the temporal and spatial analysis of the QT interval in healthy subjects and in patients with coronary artery disease (CAD) using magnetocardiography (MCG) and surface ECG. Standard 12-lead ECG and 37-channel MCG were performed in 20 healthy subjects, 23 patients with CAD without prior myocardial infarction (MI), 31 MI patients and 11 MI patients with ventricular tachycardia (VT). QTd was increased in CAD without MI compared to normals (ECG46.1 ± 6.0vs42.8 ± 5.0, P < 0.05; MCG66.8 ± 20.3vs49.7 ± 10.8, P < 0.01) and in VT compared to MI (ECG66.8 ± 16.5vs51.9 ± 16.6, P < 0.05; MCG93.6 ± 29.6vs66.8 ± 20.8, P < 0.005). In MCG, spatial distribution of QT intervals in patient groups differed from those in healthy subjects in three ways: (1) greater dispersion, (2) greater local variability, and (3) a change in overall pattern. This was quantified on the basis of smoothness indexes (SI). Normalized SI was higher in CAD without MI compared to normals (3.8 ± 1.1vs2.7 ± 0.6, P < 0.001) and in VT compared to MI (6.4 ± 1.6vs4.2 ± 1.4, P < 0.0005). For the normal-CAD comparison a sensitivity of 74% and a specificity of 80% was obtained, for MI-VT, 100% and 77%, respectively. The results suggest that examining the spatial interlead variability in multichannel MCG may aid in the initial identification of CAD patients with unimpaired left ventricular function and the identification of post-MI patients with augmented risk for VT. (PACE 2003; 26:1706,1714) [source]


Temporal Changes in the Endocardial ST Segment During the Evolution of Myocardial Infarction in Dogs

PACING AND CLINICAL ELECTROPHYSIOLOGY, Issue 11 2002
JONATHAN LESSICK
LESSICK, J., et al.: Temporal Changes in the Endocardial ST Segment During the Evolution of Myocardial Infarction in Dogs. Acute coronary occlusion causes ST-segment elevation on the body surface ECG and on the epicardial electrogram in the territory supplied by that artery. The occurrence and significance of endocardial ST changes have not been studied. The NOGA electromechanical mapping was performed on eight anesthetized dogs at baseline, immediately after occlusion of the LAD, and again at 5 hours to assess regional changes in the ST segment. At 3 days and 4 weeks the ventricles were remapped for comparison. Regional unipolar ST-segment elevation was measured for each zone from NOGA maps at 0, 80, and 120 ms after the J point. ST segments rose immediately in the infarct zones, as demarcated by echocardiography, compared to remote zones, but by 3 days had dropped below, and at 4 weeks returned to baseline values. Immediately postocclusion, ST elevation at 120 ms best differentiated between normal versus abnormal echo scores (concordance = 0.80), probably by correcting for pressure induced ST elevation. In conclusion, acute endocardial ST-segment changes occur in the infarct zone in the dog, showing a distinctive temporal evolution. [source]


Clinical Value of Electrocardiographic Parameters in Genotyped Individuals with Familial Long QT Syndrome

PACING AND CLINICAL ELECTROPHYSIOLOGY, Issue 4 2001
GEROLD MOENNIG
MOENNIG, G., et al.: Clinical Value of Electrocardiographic Parameters in Genotyped Individuals with Familial Long QT Syndrome. Rate corrected QT interval (QTc) and QT dispersion (QTd) have been suggested as markers of an increased propensity to arrhythmic events and efficacy of therapy in patients with long QT syndrome (LQTS). To evaluate whether QTc and QTd correlate to genetic status and clinical symptoms in LQTS patients and their relatives, ECGs of 116 genotyped individuals were analyzed. JTc and QTc were longest in symptomatic patients (n = 28). Both QTd and JTd were significantly higher in symptomatic patients than in asymptomatic (n = 29) or unaffected family members (n = 59). The product of QTd/JTd and QTc/JTc was significantly different among all three groups. Both dispersion and product put additional and independent power on identification of mutation carriers when adjusted for sex and age in a logistic regression analysis. Thus, symptomatic patients with LQTS show marked inhomogenity of repolarization in the surface ECG. QT dispersion and QT product might be helpful in finding LQTS mutation carriers and might serve as additional ECG tools to identify asymptomatic LQTS patients. [source]


Electrophysiologicai Characteristics of the Atrium in Sinus Node Dysfunction With and Without Postpacing Atrial Fihriliation

PACING AND CLINICAL ELECTROPHYSIOLOGY, Issue 3 2000
ANTONIO DE SISTI
DE SISTI, A., ET AL.: Electrophysiologicai Characteristics of the Atrium in Sinus Node Dysfunction With and Without Postpacing Atrial Fibrillation . In patients with sinus node dysfunction (SND) with or without associated paroxysmal atrial fibrillation (AF), the effectiveness of atrial pacing in reducing the incidence of AF is not definitive. In addition, despite several studies involving large populations of implanted patients, little attention has been paid to the electrophysioiogicai (EP) atrial substrate and the effect of permanent atrial pacing. The aim of this study is to correlate EP data and the risk of AF after DDD device implantation. We reviewed FP data of 38 consecutive patients with SND. mean age 70 ± 8 years, who were investigated free of antiarrhythmic treatment, for the evaluation of the atrial substrate. We also considered as control group 25 subjects, mean age 63 ± 14 years, referred to our EP laboratory for unexplained syncope or various atrioventricular disturbances. Following pharmacological washout and at a drive cycle length of 600 ms. effective and functional refractory periods (ERP, FRP), Sl-Al and S2-A2 latency, Al and A2 conduction duration, and latent vulnerability index (EHP/A2) were measured. AF induction was tested with up to three extrastimuli at paced cycle lengths of 600 and 400 ms in 20 patients. Induction of sustained AF (> 30 seconds) was considered as the endpoint. P wave duration on the surface ECG in lead II/Vl was also measured. DDD pacing mode was chosen in all patients with the minimal atrial rate programmed between 60 and 75 beats/min (mean 64 ± 4 beats/min). After implantation, the patients were followed-up for 29 ± 17 months and clinically documented occurrence of AF was determined. When comparing patients with SND and subjects of the control group, we did not find any significant statistical differences in terms of ERP (237 ± 33 vs 250 ± 29 ms), FRP (276 ± 30 vs 280 ± 32 ms) and Sl-Al (39 ± 16 vs 33 ± 11 ms) and S2-A2 latency (69 ± 24 vs 63 ± 25 ms). In contrast, we observed significant differences regarding Al (55 ± 19 vs 39 ± 13 ms; P < 0.001), A2 (95 ± 34 vs 57 ± 18 ms; P < 0.001) and P wave duration (104 ± 18 vs 94 ± 15 ms; P < 0.05), and ERP/A2 (2.8 ± 1.2 vs 4.8 ± 1.6; P < 0.001). When comparing patients with (n = 11) or without (n =27) postpacing AF occurrence, we did not find any difference with reference to ERP, FRP. Sl-Al, S2-A2, Al duration, or follow-up duration. In patients with postpacing AF occurrence, A2 was longer (116 ± 41 vs 87 ± 27 ms; P < 0.01), FRP/A2 lower (2.1 ± 0.4 vs 3.1 ± 1.4; P < 0.05), P wave more prolonged (116 ± 22 vs 99 ± 14 ms; P < 0.01), and preexisting AF history predominant (6/11 vs 5/27 patients; P < 0.05). No difference was observed between patients with (n = 8) and without (n = 12) AF induction during the EP study. In patients with SND, the atrial refractoriness appears normal and the most important abnormality concerns conduction slowing disturbances. Persistence of AF despite pacing stresses the importance of mechanisms responsible for AF not entirely brady-dependent. In this setting, more prolonged atrial conduction disturbances, responsible for a low vulnerability index, and a preexisting history of AF enable us to identify a high risk patient group for AF in the follow-up. sinus node dysfunction, atrial fibrillation, electrophysiologicai study, atrial pacing [source]


Abnormal P-Wave Morphology Is a Predictor of Atrial Fibrillation Development and Cardiac Death in MADIT II Patients

ANNALS OF NONINVASIVE ELECTROCARDIOLOGY, Issue 1 2010
Fredrik Holmqvist M.D., Ph.D.
Background: Several ECG-based approaches have been shown to add value when risk-stratifying patients with congestive heart failure, but little attention has been paid to the prognostic value of abnormal atrial depolarization in this context. The aim of this study was to noninvasively analyze the atrial depolarization phase to identify markers associated with increased risk of mortality, deterioration of heart failure, and development of atrial fibrillation (AF) in a high-risk population with advanced congestive heart failure and a history of acute myocardial infarction. Methods: Patients included in the Multicenter Automatic Defibrillator Implantation Trial II (MADIT II) with sinus rhythm at baseline were studied (n = 802). Unfiltered and band-pass filtered signal-averaged P waves were analyzed to determine orthogonal P-wave morphology (prespecified types 1, 2, and 3/atypical), P-wave duration, and RMS20. The association between P-wave parameters and data on the clinical course and cardiac events during a mean follow-up of 20 months was analyzed. Results: P-wave duration was 139 ± 23 ms and the RMS20 was 1.9 ± 1.1 ,V. None of these parameters was significantly associated with poor cardiac outcome or AF development. After adjustment for clinical covariates, abnormal P-wave morphology was found to be independently predictive of nonsudden cardiac death (HR 2.66; 95% CI 1.41,5.04, P = 0.0027) and AF development (HR 1.75; 95% CI 1.10,2.79, P = 0.019). Conclusion: Abnormalities in P-wave morphology recorded from orthogonal leads in surface ECG are independently predictive of increased risk of nonsudden cardiac death and AF development in MADIT II patients. Ann Noninvasive Electrocardiol 2010;15(1):63,72 [source]


Correlation of Mechanical Dyssynchrony with QRS Duration Measured by Signal-Averaged Electrocardiography

ANNALS OF NONINVASIVE ELECTROCARDIOLOGY, Issue 3 2009
F.E.S.C., George K. Andrikopoulos M.D.
Background: Preimplantation left ventricular dyssynchrony is considered a prerequisite for a beneficial response to cardiac resynchronization therapy (CRT). However, electrical dyssynchrony estimated by QRS duration (QRSd) on ECG has not been proven to be an optimal surrogate of mechanical dyssynchrony. We evaluated the correlation of mechanical dyssynchrony with QRSd as measured by signal-averaged electrocardiography (SAECG) in comparison with measurements based on conventional surface ECG and with onscreen measurements based on digital ECG. Methods: We included 49 consecutive patients with decompensated heart failure (40 men, aged 66.8 ± 9.5 years), New York Heart Association (NYHA) class II,IV, and LVEF , 40%. QRSd was calculated by manual measurement of 12-lead ECG, on-screen measurement of computer-based ECG, and calculation of total ventricular activation time on SAECG. Results: Only 60.4% of the studied patients had QRS , 120 ms based on measurements derived by SAECG compared to 69.4% by using on-screen measurement of computer-based ECG and 73.5% based on surface ECG (P = 0.041). Interventricular but not intraventricular delay was correlated with QRSd. The correlation of interventricular dyssynchrony with QRSd was stronger when measured by SAECG than by surface ECG (r = 0.45, P = 0.001 vs r = 0.35, P < 0.01). Among patients with ischemic cardiomyopathy, no significant correlation was demonstrated between mechanical dyssynchrony and QRSd. In nonischemic patients, interventricular delay was significantly correlated with QRSd measured by surface ECG (r = 0.45, P < 0.05) and SAECG (r = 0.46, P < 0.05). Conclusions: The use of SAECG results in different patient classification in wide QRS complex category as compared to surface ECG. Furthermore, QRSd measured by SAECG is correlated with interventricular but not intraventricular dyssynchrony in heart failure patients. [source]


Pilot Study: Noninvasive Monitoring of Oral Flecainide's Effects on Atrial Electrophysiology during Persistent Human Atrial Fibrillation Using the Surface Electrocardiogram

ANNALS OF NONINVASIVE ELECTROCARDIOLOGY, Issue 2 2005
Daniela Husser M.D.
Background: The relation between flecainide's plasma level and its influence on human atrial electrophysiology during acute and maintenance therapy of atrial fibrillation (AF) is unknown. Therefore, this study determined flecainide plasma levels and atrial fibrillatory rate obtained from the surface ECG during initiation and early maintenance of oral flecainide in patients with persistent lone AF and assessed their relationship. Methods and Results: In 10 patients (5 males, mean age 63 ± 14 years, left atrial diameter 46 ± 3 mm) with persistent lone AF, flecainide was administered as a single oral bolus (day 1) followed by 200,400 mg/day (days 2,5). The initial 300 mg flecainide bolus resulted in therapeutic plasma levels in all patients (range 288,629 ng/ml) with no side effects. Flecainide plasma levels increased on day 3 and remained stable thereafter. Day 5 plasma levels were lower (508 ± 135 vs 974 ± 276 ng/ml, P = 0.009) in patients with daily mean flecainide doses of 200 mg compared to patients with higher maintenance doses. Fibrillatory rate obtained from the surface electrocardiogram measuring 378 ± 17 fpm at baseline was reduced to 270 ± 18 fpm (P < 0.001) after the flecainide bolus but remained stable thereafter. Fibrillatory rate reduction was independent of flecainide plasma levels or clinical variables. Conclusion: A 300 mg oral flecainide bolus is associated with electrophysiologic effects that are not increased during early maintenance therapy in persistent human lone AF. In contrast to drug plasma levels, serial analysis of fibrillatory rate allows monitoring of individual drug effects on atrial electrophysiology. [source]


R-Wave Detection by Subcutaneous ECG.

ANNALS OF NONINVASIVE ELECTROCARDIOLOGY, Issue 1 2001
Possible Use for Analyzing R-R Variability
Background: Atrial arrhythmia (AA) discrimination remains a technological challenge for implanted cardiac devices. We examined the feasibility of R-wave detection by a subcutaneous far field ECG (SFFECG) and analysis of these signals for R to R variability as an indicator of atrial arrhythmia (AA). Methods: Surface ECG and SFFECG (from the pacemaker pocket) were recorded in sixteen patients (61.5 ± 11.4 years) with AA. The SFFECG was recorded with a pacemaker sized four electrode array acutely placed in the pacemaker pocket during implantation. The signals were analyzed to obtain peak-to-peak R wave amplitude and R to R interval variability (indicative of AAs). Results: In sixteen patients R waves were visually discernible in all recordings. The percentage over and under detection for automatic R wave recognition SFFECG was 3 and 9%, respectively. R to R variability analysis using the SFFECG produced results concordant to those using the surface ECG. Conclusion: SFFECG might be a helpful adjunct in implantable device systems for detection of R waves and may be used for measurement of R to R variability. A.N.E. 2001;6(1):18,23 [source]


Electrocardiographic Quantitation of Heterogeneity of Ventricular Repolarization

ANNALS OF NONINVASIVE ELECTROCARDIOLOGY, Issue 1 2000
Peter M. Okin M.D.
Background:QT interval dispersion (QTd) measured from the surface ECG has emerged as the most common noninvasive method for assessing heterogeneity of ventricular repolarization. Although QTd correlates with dispersion of monophasic action potential duration at 90% repolarization and with dispersion of recovery time recorded from the epicardium, total T-wave area, representing a summation of vectors during this time interval, has been shown to have the highest correlation with these invasive measures of dispersion of repolarization. However, recent clinical studies suggest that the ratio of the second to first eigenvalues of the spatial T-wave vector using principal component analysis (PCA ratio) may more accurately reflect heterogeneity of ventricular repolarization. Methods:To better characterize the ECG correlates of surface ECG measures of heterogeneity of ventricular repolarization and to establish normal values of these criteria using an automated measurement method, the relations of QRS onset to T-wave offset (QTod) and to T-wave peak (QTpd) dispersion and the PCA ratio to T-wave area and amplitude, heart rate, QRS axis and duration, and the QTo interval were examined in 163 asymptomatic subjects with normal resting ECGs and normal left ventricular mass and function. QTod and QTpd were measured by computer from digitized ECGs as the difference between the maximum and minimum QTo and QTp intervals, respectively. Results:In univariate analyses, a significant correlation was found between the sum of the T-wave area and the PCA ratio (R =,0.46, P < 0.001), but there was no significant correlation of the sum of T-wave area with QTod (R = 0.11, P = NS) or QTpd (R=0.09, P = NS). There were only modest correlations between QTod and QTpd (R = 0.45) and between the PCA ratio and QTod (R = 0.29) and QTpd (R = 0.49) (each P < 0.001). In stepwise multivariate linear regression analyses, the PCA ratio was significantly related to the sum of T-wave area, T-wave amplitude in aVL, and to female gender (overall R = 0.54, P < 0.001), QTod correlated only with the maximum QTo0 interval (R = 0.39, P < 0.001), and QTpd was related to heart rate and QRS axis (overall R = 0.36, P <0.001). In addition, the normal interlead dispersion of repolarization as measured by QTod was significantly greater than dispersion measured by QTod (23.5 ± 11.5 ms vs 18.3 ± 11.2 ms, P < 0.001). Conclusions: These findings provide new information on ECG measures of heterogeneity of repolarization in normal subjects, with a significantly higher intrinsic variability of Q to T-peak than Q to T-offset dispersion and only modest correlation between these wo measures. The independent relation of the PCA ratio to the sum of T-wave area suggests that the PCA ratio may be a more accurate surface ECG reflection of the heterogeneity of ventricular repolarizat on. A.N.E. 2000;5(1):79,87 [source]


The comparative effects of telmisartan and ramipril on P-wave dispersion in hypertensive patients: A randomized clinical study

CLINICAL CARDIOLOGY, Issue 6 2005
Turgay Celik M.D.
Abstract Background: Prolongation of P-wave times and increase of P-wave dispersion (PWD) were shown to be independent predictors of atrial fibrillation (AF). Angiotensin II receptor blockers (AARBs) and angiotensin-converting enzyme inhibitors (ACEIs) have beneficial effects on atrial conduction times. However, there are not enough data about the comparative effects of those drugs onPWD. Hypothesis: We aimed to compare the effects of telmisartan and ramipril on PWD after 6-month treatment in hypertensive patients. Methods: In all, 100 newly diagnosed hypertensive patients were enrolled in the study and were randomly assigned to two groups. Group 1 and Group 2 each consisted of 50 patients, taking daily doses of 80 mg telmisartan and 10 mg ramipril, respectively. Twelve-lead surface electrocardiograms (ECG) were recorded from all patients before and after 6-month drug therapy. The P-wave duration (Pdur) measurements were calculated from the 12-lead surface ECG. Results: When pretreatment PWD and P maximum values were compared with post-treatment values, a statistically significant decrease was found in both groups after 6 months (Group 1 and 2; p < 0.001 for PWD and Pmaximum). P-wave dispersion and Pmaximum values after treatment in Group 1 were statistically significantly lower than those in Group 2 after the 6-month treatment period (p = 0.01 for PWD; p = 0.008 for Pmaximum). Conclusions: Telmisartan has a much greater lowering effect on PWD and Pmaximum values than ramipril. This finding may be important in the prevention of AF in hypertensive patients. [source]


Autonomic nervous system functions in children with breath-holding spells and effects of iron deficiency

ACTA PAEDIATRICA, Issue 9 2005
Abdülkerim Kolkiran
Abstract Aim: To analyse the activity of the autonomic nervous system during breath-holding spells, we assessed the ECG changes, including ventricular repolarization parameters before and during the spell. We also analysed the effects of iron deficiency on these ECG parameters. Methods: The study group consisted of 37 children with breath-holding spells (30 cyanotic, 7 pallid) (mean age±SD: 12.9±10.8 mo). Twenty-six healthy children (mean age±SD: 14.4±8.6 mo) served as a control group. All patients and controls had standard 12-lead simultaneous surface ECG. All patients had ECG recordings during at least one severe breath-holding spell obtained by "event recorder". Traces obtained by "event recorder" were analysed in terms of mean heart rate and the frequency and duration of asystole during the spell. Results: Respiratory sinus arrhythmia on standard ECGs and asystole frequency during spells were higher in patients with pallid breath-holding spells. Patients with iron deficiency had a lower frequency of respiratory sinus arrhythmia and prolonged asystole time during the spell. There was no difference in terms of ventricular repolarization parameters (QT/QTc intervals and QT/QTc dispersions) between patients and controls and between patient subgroups (cyanotic versus pallid). Conclusion: These results confirmed the presence of autonomic dysregulation in children with breath-holding spells. Iron deficiency may have an impact on this autonomic dysregulation. Ventricular repolarization was unaffected in patients with breath-holding spells. [source]


Mechanism of Repolarization Change During Initiation of Supraventricular Tachycardia

JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, Issue 11 2004
YENN-JIANG LIN M.D.
Introduction: Previous literature has documented the association between narrow QRS supraventricular tachycardia (SVT) and pronounced ST-T segment change. The aim of this study was to evaluate repolarization changes during SVT initiation and demonstrate the possible mechanism. Methods and Results: Fifty-one consecutive patients (20 men and 31 women; mean age 46.1 ± 16.4 years) with narrow QRS SVT (32 patients with AV nodal reentrant tachycardia and 19 patients with AV reentrant tachycardia) were included. We retrospectively analyzed the intracardiac recordings and ST-T segment changes on 12-lead surface ECGs during SVT initiation. Twenty-six (51%) patients developed ST segment repolarization changes during SVT initiation. Patients with shorter baseline sinus cycle length, shorter tachycardia cycle length, elevated systolic blood pressure before tachycardia induction, and greater reduction of systolic blood pressure had a higher incidence of repolarization changes. However, multivariate analysis showed that reduction of systolic blood pressure after SVT induction was the only independent predictor of repolarization changes. Furthermore, the maximal degree of ST segment depression during SVT correlated with the reduction of systolic blood pressure (r = 0.75, P < 0.001). Conclusion: Repolarization changes during SVT initiation were caused mainly by concurrent hemodynamic change after SVT initiation with abrupt cycle length shortening. [source]