Load Cell (load + cell)

Distribution by Scientific Domains


Selected Abstracts


DEVELOPMENT OF A BINOCULAR TYPE SIX-COMPONENT LOAD CELL

EXPERIMENTAL TECHNIQUES, Issue 1 2003
D.I. Kang
First page of article [source]


The influence of impact object characteristics on impact force and force absorption by mouthguard material

DENTAL TRAUMATOLOGY, Issue 1 2004
Tomotaka Takeda
Abstract,,, Most impact force and impact energy absorption tests for mouthguards have used a steel ball in a drop-ball or the pendulum device. However, in reality most sports-related trauma is caused by objects other than the steel ball, e.g. various sized balls, hockey puck, or bat or stick. Also, the elasticity, the velocity and the mass of the object could change the degree and the extent of injuries. In this study, we attempted to measure the impact force from actual sports equipment in order to clarify the exact mechanism of dental-related sports injuries and the protective effects of mouthguards. The present study was conducted using the pendulum impact device and load cell. Impact objects were removable. Seven mobile impact objects were selected for testing: a steel ball, baseball, softball, field hockey ball, ice hockey puck, cricket ball, and wooden baseball bat. The mouthguard material used in this study was a 3-mm-thick Drufosoft (Dreve-Dentamid GmbH, Unna, Germany), and test samples were made of the one-layer type. The peak transmitted forces without mouthguard ranged from the smallest (ice hockey stick, 46.9 kgf) to the biggest (steel ball, 481.6 kgf). The peak transmitted forces were smaller when the mouthguard was attached than without it for all impact materials but the effect was significantly influenced by the object type. The steel ball showed the biggest (62.1%) absorption ability while the wooden bat showed the second biggest (38.3%). The other balls or the puck showed from 0.6 to 6.0% absorbency. These results show that it is important to test the effectiveness of mouthguards on specific types of sports equipment. In future, we may select different materials and mouthguard designs suitable for specific sports. [source]


Viscous behaviour of dry sand

INTERNATIONAL JOURNAL FOR NUMERICAL AND ANALYTICAL METHODS IN GEOMECHANICS, Issue 15 2007
D. Pham Van Bang
Abstract Tests on air-dried Hostun sand were performed on a newly developed triaxial apparatus, which allows static and dynamic measurements at ,Département Génie Civil et Bâtiment' (DGCB) of ENTPE. The prototype was designed to perform tests in order to investigate the viscous effects of sand from small strain (some 10,5m/m) up to intermediate strain (some 10,3m/m). The accuracy of the measuring system is ensured by the use of local non-contact displacements transducers and by the use of an internal load cell. The viscous behaviour of sand is experimentally studied for loose and dense specimens for confining pressure values from 80 up to 400 kPa and considering different histories of loading. Three types of viscous loading: creep periods; stress relaxation periods; and paths with stepwise changes in the strain rate are specifically analysed. The proposed viscous modelling is issued from a three-component general framework and is able to reproduce the experimental observations. More specifically, the peculiar behaviour of sand observed during a stepwise increase or decrease in the strain rate is detailed and modelled by the viscous evanescent (VE) model, developed at DGCB/ENTPE. Based on the three types of viscous tests, a simple relation is finally proposed for the viscous parameters of the VE model. Comparisons between data and simulations validate the approach. Copyright © 2007 John Wiley & Sons, Ltd. [source]


The Response of Skin to Applied Stress: Investigation of Bitemark Distortion in a Cadaver Model,

JOURNAL OF FORENSIC SCIENCES, Issue 1 2010
Mary A. Bush D.D.S.
Abstract:, Knowledge of distortional properties of skin is important in bitemark analysis. Thus, the response of skin to stress from bites was investigated. Four sets of models were created from the dentition of one individual. Anterior teeth were systematically removed to vary contact surface area. A biting apparatus was constructed with an integrated load cell. Forty-six bites were created perpendicular to Langer lines on six cadavers. Rate of force application and bite pressure were controlled. Metric/angular measurement and hollow volume overlays were employed. Distortion produced by each dentition was calculated and assessed. Results showed that as teeth impressed loose tissue, mesial/distal distance increased, angles of rotation flattened, and inter-canine distance lengthened. An opposite effect was seen in tight tissue. When the surface area of the dentition was reduced, a mixture of these effects was observed. Conclusions indicated that stiffness of the tissue was the most important variable in bitemark distortion. [source]


Impact of gravity loading on post-stroke reaching and its relationship to weakness

MUSCLE AND NERVE, Issue 2 2007
Randall F. Beer PhD
Abstract The ability to extend the elbow following stroke depends on the magnitude and direction of torques acting at the shoulder. The mechanisms underlying this link remain unclear. The purpose of this study was to evaluate whether the effects of shoulder loading on elbow function were related to weakness or its distribution in the paretic limb. Ten subjects with longstanding hemiparesis performed movements with the arm either passively supported against gravity by an air bearing, or by activation of shoulder muscles. Isometric maximum voluntary torques at the elbow and shoulder were measured using a load cell. The speed and range of elbow extension movements were negatively impacted by actively supporting the paretic limb against gravity. However, the effects of gravity loading were not related to proximal weakness or abnormalities in the elbow flexor,extensor strength balance. The findings support the existence of abnormal descending motor commands that constrain the ability of stroke survivors to generate elbow extension torque in combination with abduction torque at the shoulder. Muscle Nerve, 2007 [source]


Motor block in regional anaesthesia

ANAESTHESIA, Issue 12 2001
apparatus
A device based on a load cell was constructed to measure the strength of foot dorsiflexion and plantarflexion. Performance of the device was evaluated for both movements. The influence of foot position within the device, its use over a 30-min period at 30-s intervals and the effect of the removal and reapplication of the device on measured force of dorsiflexion and plantarflexion was studied in six volunteers. Both dorsiflexion and plantarflexion are suitable movements on which to base a device to quantify the density of motor block during the onset and offset of neuraxial block. Dorsiflexion has a number of advantages: muscle strength is independent of knee position, and therefore a below-knee device can be constructed; strength of dorsiflexion is less affected by the foot position; we found the device easier to apply using dorsiflexion as the heel tended to self-locate; innervation of the muscles responsible for dorsiflexion involves fewer spinal segments. [source]


Assessment of Finger Forces and Wrist Torques for Functional Grasp Using New Multichannel Textile Neuroprostheses

ARTIFICIAL ORGANS, Issue 8 2008
Marc Lawrence
Abstract:, New multichannel textile neuroprotheses were developed, which comprise multiple sets of transcutaneous electrode arrays and connecting wires embroidered into a fabric layer. The electrode arrays were placed on the forearm above the extrinsic finger flexors and extensors. Activation regions for selective finger flexion and wrist extension were configured by switching a subset of the array elements between cathode, anode, and off states. We present a new isometric measurement system for the assessment of finger forces and wrist torques generated using the new neuroprostheses. Finger forces (from the middle phalanxes) were recorded using five load cells mounted on a "grasp handle" that can be arbitrarily positioned in space. The hand and the grasp handle were rigidly mounted to a 6-degree of freedom load cell, and the forces and torques about the wrist were recorded. A vacuum cushion was used to comfortably fixate the forearm. The position and orientation of the forearm, wrist, fingers, and handle were recorded using a new three-dimensional position measurement system (accuracy <±1 mm). The measurement system was integrated into the real-time multichannel transcutaneous electrode environment, which is able to control the spatiotemporal position of multiple activation regions. Using the combined system and textile neuroprosthesis, we were able to optimize the activation regions to produce selective finger and wrist articulation, enabling improved functional grasp. [source]


Overcoming Abnormal Joint Torque Patterns in Paretic Upper Extremities Using Triceps Stimulation

ARTIFICIAL ORGANS, Issue 3 2005
Thierry Keller
Abstract:, The goal of this research project was to quantitatively assess whether transcutaneous triceps stimulation can overcome the expression of abnormal torque patterns in the paretic upper limb of subjects with hemiparetic stroke. Abnormal torque patterns consist of strong coupling between shoulder abduction (SAB) and elbow flexion (EF) or between elbow extension (EE) and shoulder adduction (SAD) torques. Both patterns reduce the active range of motion during arm movements. Eight chronic stroke subjects with moderate to severe (Fugl-Meyer assessment scores of 21/66,36/66) upper limb motor impairment participated in this study. Shoulder and elbow joint torques were measured with a 6-degrees-of-freedom load cell under isometric conditions, while the triceps muscle was stimulated to generate EE torques. At the same time the subjects were asked to lift up their arm to generate different SAB torque levels. The obtained isometric results showed that electrical stimulation can overcome abnormal torque patterns in chronic stroke subjects while generating SAB. This is likely to have potential benefits to increase the, reaching, workspace, of, the, paretic, arm. [source]


Shear bond strength of luting agents to fixed prosthodontic restorative core materials

AUSTRALIAN DENTAL JOURNAL, Issue 4 2009
N Capa
Abstract Background:, Bonding properties of luting cements are important for retention of restorative core materials. The aim of this study was to compare the bonding performance of a resin-modified glass ionomer cement and a self-adhesive resin cement to various fixed prosthodontic core materials. Methods:, Cylindrical specimens with a thickness of 2 mm and a diameter of 5 mm were fabricated from Au-Pd-Ag, Co-Cr, Ni-Cr-Mo, Ni-Cr-Fe alloys, titanium, zirconia and Empress II (n = 20). Each group was divided into two subgroups to be luted with two different luting agents. Composite resin blocks were cemented onto specimens with RelyXUnicem and FujiCem. A shear bond strength machine with 50 kg load cell and 0.50 mm/min crosshead speed was used. Kruskal Wallis test, Dunn's Multiple Range test and Mann-Whitney-U test were used for statistical analysis. The results were evaluated in a confidence interval of p < 0.05. Results:, The highest bond strength was obtained between Ni-Cr-Fe-RelyXUnicem (8.22 ± 2.15 MPa) and the lowest was between Empress II-FujiCem (1.48 ± 0.9 MPa). In FujiCem groups, Co-Cr and Ni-Cr-Fe showed significantly higher bond strength than Au-Pd-Ag and Empress II. In RelyX Unicem groups, Ni-Cr-Fe showed higher bond strength than Empress II. Conclusions:, The types of luting agents and restorative core materials may have a significant influence on bond strength. [source]


Comparison of girth materials, girth tensions and their effects on performance in racehorses

AUSTRALIAN VETERINARY JOURNAL, Issue 1-2 2005
J BOWERS
Objective To compare the effect of girth materials and commonly used girth tensions on athletic performance of racehorses and to test the length tension properties of commercially available girths. Procedure Seven horses were exercised at speeds to produce 95% of maximal heart rates on 15 occasions using a randomised block design, and girthed with 5 different girths at 3 nominal tensions of 6, 12 or 18 kg. The girths used were a standard elastic race girth, an ,American' elastic race girth, an elastic race girth twice the normal width, a standard canvas race girth and a canvas race girth at twice the normal width. Tension in the girth was recorded continuously using an in-line load cell connected to a physiograph. Horses ran to fatigue on a treadmill inclined at 10% slope. Tensions were measured at peak inhalation (T/inh) and exhalation (T/exh), recorded at rest (rest) and during exercise (ex). An analysis of variance was used to compare the mean run to fatigue times (RTFT) between girth types and tensions, multiple pair-wise comparisons were then carried out using Tukey's test where significant differences were found. The length-tension relationships of five commercially available girths for training and racing of Thoroughbred racehorses were studied by the application of standardized weights in series to multiple samples of each type of girth. Measurements were taken in a controlled environment and analysis of variance was used to compare the means for length-tension of each girth type. Results The elastic and the ,American' elastic girths produced significantly longer RTFT when compared to the standard canvas girth (P=0.01 and P = 0.001 respectively). Also girths tensioned at Texhrest 6 kg and Texhrest 12 kg produced significantly longer RTFT than when girthed at Texhrest 18 kg (P=0.03 and P = 0.08 respectively). There were significant differences between the commercially available girth types at each tension (P < 0.05), but differences were not significant between girths of the same type. Girths with an elastic component reached their peak for maximum extension at 14.5 kg and thereafter their extension declined. Conclusion The type of girth and the tension at which it is applied affects athletic performance. Lower girth tensions and the use of elastic materials in the girth would appear to optimise performance. However according to this study and our previous study, none of the commercially available girths studied would adequately protect against the potentially detrimental effects of overtightening on athletic performance. [source]


Tensions used on girths on Thoroughbred racehorses

AUSTRALIAN VETERINARY JOURNAL, Issue 8 2000
J BOWERS
Objective To determine commonly used girth tensions in Thoroughbred racehorses in the Melbourne metropolitan area. Design A prospective industry survey. Methods Strappers were instructed to tension girths, using the same saddle, under-saddle packing and girth, sufficiently to hold the saddle for racing. These tensions were continuously recorded by an in-line load cell in the girth. Seventeen strappers from five stables participated in the study, which was conducted on 91 horses. Results Resting girth tensions varied considerably. The mean tension on inhalation was 13 ± 0.4 kg. Male strappers recorded higher tensions than females for all indices measured. Age and girth size of the horse did not significantly influence results. Conclusion This study confirms that there is no standard for application of a girth and saddle in the Thoroughbred racing industry and the range of tensions applied was large. Strappers were unable to reliably apply the same tensions between horses. Values recorded at the extremes of the range seem inappropriate for racing conditions as low values would result in slippage of the saddle and higher values may result in girth galls and possible reduction in athletic performance. Further work is necessary to determine the point at which girth tension affects performance. [source]


4336: Acute mechanical effect of riboflavin corneal cross-linking

ACTA OPHTHALMOLOGICA, Issue 2010
A PENNOS
Purpose To study the acute mechanical effect of corneal cross-linking in vitro. Methods Twenty porcine eyes were harvested from a local abattoire. The epithelium was removed using a rotating brush and corneal buttons having diameter of 8 mm were excised using a Barron's PKP trephine. The buttons were left immersed for four hours in Dextran solution (20%W/V) in order to regulate their hydration. The buttons were immersed in riboflafin 0.1% in dextran 20% and separated in two groups (10 buttons each) where one group was irradiated for 30 minutes with UV light while the other group was not irradiated and served as control. After treatment mechanical measurements were performed by means of a purposely-developed device featuring a stepping motor and a load cell. The motor shaft has a spherical tip that compresses the corneal button against a flat surface mounted on the load cell. The slope of force versus normalized deformation was recorded for each specimen. Results The control group had an average slope of 3.55 (± 1.81) while the irradiated group had 4.47 (±1.80). This difference was not statistically significant. (p=0.27) Conclusion This compressibility test did not reveal statistically significant changes between the two groups immediately after UV corneal cross-linking. It remains to be elucidated whether different tests (e.g. elongation) may yield different results. [source]


Comparison of forces transmitted through different EVA mouthguards

DENTAL TRAUMATOLOGY, Issue 4 2006
Craig F. Duhaime
Abstract,,, Athletic mouthguards have been recommended for decades with varying levels of athlete acceptance. Issues related to compliance center around the ability to breath and speak while wearing the mouthguards. Fabrication techniques have changed over time to a two-layer ethylene vinyl acetate mouthguard fabricated on a high-pressure machine. The reported ideal thickness of these mouthguards has been somewhat variable depending on the sport and anticipated level of risk. Recent research however, has identified 4 mm as the optimal thickness of EVA. In this study an acrylic dental cast was fabricated and mounted to a drop impact fixture. Mouthguards of varying ply, thickness and palatal coverage were fabricated and tested in the fixture. Strain gauges and load cells were used to evaluate the effect of ply, thickness, and palatal coverage on the ability of these mouthguards to minimize transmitted forces. The purpose of this study was to identify those variables of mouthguard construction that will minimize the overall transmitted force of impact to the anterior dentition. [source]


Mass and size distribution of firebrands generated from burning Korean pine (Pinus koraiensis) trees

FIRE AND MATERIALS, Issue 1 2009
Samuel L. Manzello
Abstract The present study reports on a series of real-scale fire experiments that were performed to determine the mass and size distribution of firebrands generated from Korean Pine (Pinus koraiensis), a common conifer species indigenous to China, Japan, and Korea. The experiments were performed at the Building Research Institute in Tsukuba, Japan. The total tree height was fixed at 4.0,m and tree moisture content was varied to examine the influence that this parameter has on the mass and size distribution of the firebrands that are produced, under ambient wind conditions. The firebrands were collected using an array of pans. The pans used for firebrand collection were filled with water. This ensured that firebrands would be quenched as soon as they made contact with the pans. The firebrands were subsequently dried and the mass and size of more than 500 firebrands were measured. The Korean pine trees were also mounted on load cells during burning to determine the temporally resolved mass loss profiles. The mass loss data were used to compare the total amount of mass collected as firebrands with the total amount of mass burned. Results of this study are presented and compared with the mass and size distribution of firebrands collected from burning Douglas-fir trees, a conifer tree species indigenous to the U.S.A. Copyright © 2008 John Wiley & Sons, Ltd. [source]


The effects of partial and total interosseous membrane transection on load sharing in the cadaver forearm

JOURNAL OF ORTHOPAEDIC RESEARCH, Issue 4 2001
Michael F. Shepard
This study was performed to examine the effects of partial and total transection of the interosseous membrane (IOM) on load transfer in the forearm. Twenty fresh frozen forearms were instrumented with custom designed load cells placed in the proximal radius and distal ulna. Simultaneous measurements of load cell forces, radial head displacement relative to the capitellum, and local tension within the central band of the IOM were made as the wrist was loaded to 134 N with the forearm at 90° of elbow flexion and in neutral pronation supination. For valgus elbow alignment (radial head contacting the capitellum), mean force carried by the distal ulna was 7.1% of the applied wrist force and mean force transferred from radius to ulna through the IOM was 4.4%. For varus elbow alignment (mean 2.0 mm gap between the radial head and capitellum), mean distal ulna force was 28% and mean IOM force was 51%. Section of the proximal and distal one-thirds of the IOM had no significant effect upon mean distal ulnar force or mean IOM force. Total IOM section significantly increased mean distal ulnar force for varus elbow alignment in all wrist positions tested. The mean level of applied wrist force necessary to close the varus gap (89 N) decreased significantly after both partial IOM section (71 N) and total IOM section (25 N). The IOM became loaded only when the radius displaced proximally relative to the ulna, closing the gap between the radius and capitellum. As the radius displaced proximally, the wrist becomes increasingly ulnar positive, which in turn leads to direct loading of the distal ulna. This shift of force to the distal ulna could present clinically as ulnar sided wrist pain or as ulnar impaction after IOM injury. © 2001 Orthopaedic Research Society. Punlished by Elsevier Science Ltd. All rights reserved. [source]


A LARGE ELECTRONIC WEIGHING LYSIMETER SYSTEM: DESIGN AND INSTALLATION,

JOURNAL OF THE AMERICAN WATER RESOURCES ASSOCIATION, Issue 4 2002
Gholam-Abbas Barani
ABSTRACT: To measure crop evapotranspiration, a large double tank, electronic weighting lysimeter system was designed and installed at the Shahid Bahonar University farm, Kerman, Iran. The system was installed in a 50 m2 underground building. It includes two tanks of 3.00 m in diameter and 1.75 m deep. The weighing mechanism for each tank is a set of three compression strain gage load cells, which are fixed on 1.20 m height column above the floor. According to the specification of the load cells, the maximum possible weighing error may be about 0.01 percent of total mass, which is equivalent to 0.28 mm of water, but the measured error was equal to 1 kg mass, which is equivalent to 0.14 mm of water. The load cell data from each tank and the on-site environmental data (temperature, humidity, and wind velocity and direction) are automatically recorded and saved in a personal computer hard disk for further use and analysis. [source]


Human ex vivo bone tissue strains around immediately loaded implants supporting maxillary overdentures

CLINICAL ORAL IMPLANTS RESEARCH, Issue 6 2005
vanç Akça
Abstract Purpose: To evaluate ex vivo bone tissue strains around maxillary implants supporting immediately loaded bar-retained overdentures. Material and methods: Ninety degree two-element rosette strain gauges were bonded on the labial bone of four ITI® dental implants placed in the maxillary anterior region of four completely edentulous maxilla of fresh human cadavers. The installation torque value (ITV) of each implant was measured by a custom-made torque wrench and resonance frequency analyses (RFAs) were undertaken. A bar-retained overdenture was fabricated for each cadaver, and two miniature load cells were integrated in the first molar region of the overdentures for controlled loading experiments. Strain measurements were performed at a sample rate of 10 kHz and under a maximum load of 100 N, simultaneously monitored from a computer connected to a data acquisition system. Finally, removal torque values (RTVs) of the implants were measured. Results: RFA values did not mirror ITVs, while RTVs of implants were slightly lower than the ITVs. Any correlation could not be obtained between RFA values and ITVs or RTVs. Maximum strains around loaded implants ranged between ,100 and ,550 ,, under 25,100 N. The axial and lateral strain values of posterior implants of both sides were higher than those of anterior implants under all loads (P<0.05). Conclusion: Because occlusal forces in humans tend to decrease because of age-related factors, maximum strains around immediately loaded implants supporting maxillary overdentures fall within physiologic levels. Résumé Le but de cette étude a été d'évaluer les tensions du tissu osseux ex vivo autour d'implants maxillaires portant des prothèses amovibles retenues sur une barre placée immédiatement après l'insertion des implants. Deux jauges de force en rosette à 90° ont été attachées au côté lingual de l'os de quatre implants dentaires ITI® placés dans la région antérieure du maxillaire de quatre cadavres humains frais aux maxillaires édentées. La valeur du couple de torsion lors de l'installation (ITV) de chaque implant a été mesurée par un couple fabriqué et des analyses de fréquence de résonnance (RFA) ont été faites. Une prothèse retenue par une barre a été fabriquée pour chaque cadavre et deux cellules de charge miniatures ont été intégrées dans la région de la première molaire de ces prothèses pour les expériences de charge contrôlées. Les mesures de force ont été effectuées à un taux d'échantillonnage de 10 kHz et sous une charge maximale de 100 N, suivies simultanément par un ordinateur connectéà un système d'acquisition de données. Finalement les valeurs des couples de torsion à l'enlèvement (RTV) des implants ont été mesurées. Les valeurs RFA n'étaient pas en ligne avec les ITV tandis que les RTV des implants étaient légèrement inférieurs aux ITV. Aucune corrélation n'a pûêtre établie entre les valeurs RFA et ITV ou RTV. Les forces maximales autour des implants chargés s'étalaient de ,100 ,e à,550 ,e sous 25 N à 100 N. Les valeurs des forces axiales et latérales des implants postérieurs des deux côtés étaient supérieures à celles des implants antérieurs sous toutes les charges (P<0.05). Zusammenfassung Ziel: Es war das Ziel dieser Arbeit, an bis vor kurzem vitalen Knochengewebe Spannungen um Oberkieferimplantate zu untersuchen, welche sofortbelastete und stegverankerte Hybridprothesen tragen. Material und Methoden: Auf den labialen Knochen von vier ITI-Implantaten in der vorderen Region des zahnlosen Oberkiefers von vier frisch verstorbenen Menschen klebte man Dehnmessinstrumente an. Man mass den Wert der Eindrehkraft (ITV) jedes einzelnen Implantates mit einem handelsblichen Drehmomentschlssel und fhrte eine Resonanzfrequenzanalyse (RFA) durch. Man stellte fr jede Leiche eine stegverankerte Hybridprothese her und man baute in der Region der ersten Molaren zwei kleine Messgerte ein, die kontrollierte Belastungsexperimente erlaubten. Dehnmessungen fhrte man mit einer Frequenz von 10 kHz und einer maximalen Belastung von 100 N durch, simultan aufgezeichnet von einem mit dem Messsystem verbundenen Komputer. Schliesslich mass man auch die Krfte, die es brauchte um die Implantate wieder auszudrehen (RTV). Resultate: Die RFA wiederspiegelte die ITV-Werte nicht, whrenddem die RTV-Werte der Implantate geringfgig tiefer waren als die ITV-Werte. Man konnte weder zwischen der RFA, den ITV-Werten oder den RTV-Werten eine Korrelation herstellen. Die maximalen Dehnungen um die Implantate erreichte Werte zwischen ,100 E und ,550 E bei einer Belastung zwischen 25 N und 100 N. Die axialen und lateralen Dehnkrfte bei posterioren Implantaten waren unter smtlichen getesteten Belastungen beidseits grsser als diejenigen der anterioren Implantate (P<0.05). Resumen Propósito: Evaluar las tensiones del tejido óseo ex vivo alrededor de implantes maxilares soportando sobredentaduras retenidas por barras con carga inmediata. Material y Métodos: Se pegaron dos indicadores de tensión de roseta de dos elementos de 90° en el hueso labial de cuatro implantes dentales ITI® colocados en la región maxilar anterior de cuatro maxilares completamente edéntulos de cadáveres humanos frescos. Se midió el valor del torque de instalación (ITV) de cada implante por medio de una chicharra de torque hecha a medida y se llevó a cabo análisis de frecuencia de resonancia (RFA). Se fabricó una sobredentadura retenida por barras para cada cadáver y se integraron dos células miniatura de carga en la región del primer molar de las sobredentaduras para los experimentos de carga controlada. Se llevaron a cabo mediciones de la tensión a un índice de muestra de 10 kHz y bajo una carga máxima de 100 N, simultáneamente monitorizada desde un ordenador conectado a sistema de adquisición de datos. Finalmente, se midieron los valores del torque de remoción (RTV) de los implantes. Resultados: Los valores de RFA no se reflejaron en los ITVs, mientras que los RTVs de los implantes fueron ligeramente más bajos que los ITVs. No se pudieron obtener correlaciones entre los valores de RFA y los ITVs o RTVE. Las tensiones máximas alrededor de los implantes cargados variaron entre ,100 ,, y ,550 ,, bajo 25 N a 100 N. Los valores de las tensiones axiales y laterales de los implantes posteriores de ambos lados fueron mayores que aquellos implantes anteriores bajo todas las cargas (P<0.05). [source]