Lead Impedance (lead + impedance)

Distribution by Scientific Domains


Selected Abstracts


Intraoperative Comparison of a Subthreshold Test Pulse with the Standard High-Energy Shock Approach for the Measurement of Defibrillation Lead Impedance

JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, Issue 1 2006
ANDREAS SCHUCHERT M.D.
There are two methods to measure shocking lead impedance: delivery of high-energy shocks that require patient sedation, and the painless measurement of impedance from subthreshold test pulses. The aim of this study was to compare the two methods. Methods: The study included 131 patients implanted with a standard DR (n = 71) or VR (n = 60) ICD connected to either single-coil (n = 39) or dual-coil (n = 92) defibrillation leads. The noninvasive high-energy impedance test was done using a 17 J shock after induction of ventricular tachyarrhythmias and compared to a 0.4 ,J test pulse used by the ICD for the subthreshold measurements. Results: Defibrillation lead impedance measurements were not significantly different between patients with the same shocking vector configuration. In patients with a single-coil defibrillation lead the impedance was 62 ± 9 , with the high-energy shock and 62 ± 8 , with the subthreshold test pulses (P = 0.13). Patients with a dual-coil configuration recorded average impedances of 40 ± 5 , from both tests (P = 0.44). While there was no difference in values recorded within each lead configuration, there was a significant difference in impedance between the single-coil and the dual-coil patient groups (P = 0.001). Conclusions: There was no significant difference between shocking lead impedances measured with the high-energy shock or the subthreshold test pulses. This offers the possibility of noninvasive, low-energy serial measurements of shocking lead impedance at follow-up visits and removing the need for sedation. [source]


Subthreshold Test Pulses Versus Low Energy Shock Delivery to Estimate High Energy Lead Impedance in Implanted Cardioverter Defibrillator Patients

PACING AND CLINICAL ELECTROPHYSIOLOGY, Issue 1p2 2003
DIRK VOLLMANN
VOLLMANN, D., et al.: Subthreshold Test Pulses Versus Low Energy Shock Delivery to Estimate High Energy Lead Impedance in Implanted Cardioverter Defibrillator Patients. The high energy lead impedance is valuable for detecting lead failure in ICDs, but until recently shock delivery was necessary for high energy impedance measurement. This study compared the use of subthreshold test pulses and low energy test shocks to estimate the high energy impedance. Immediately after implantation of Ventak Prizm ICDs in 29 patients, the lead impedance was measured with five subthreshold (0.4 ,J) test pulses, 5 low energy (1.1 J) shocks, and two to three high energy(16 ± 4.5 J)shocks. The mean impedances measured using high energy shocks, low energy shocks, and subthreshold pulses were42.0 ± 7.3 ,, 46.5 ± 8.1 ,, and42.4 ± 7.1 ,, respectively. The impedances measured using high and low energy shocks differed significantly(P <0.0001), while those obtained by high energy shocks and low energy pulses did not(P = 0.63). According to the Pearson correlation coefficient, the impedance measurements with subthreshold pulses and low energy shocks were both closely correlated(P < 0.0001)with impedance values determined with high energy shocks. However, while the impedance values tended to be higher when measured with low energy shocks, the concordance correlation coefficient (c) was higher for subthreshold test pulse versus high energy shock(c = 0.92)than for low versus high energy shock(c = 0.73). Furthermore, the intraindividual variability of impedance measurements was lower with subthreshold pulse measurements than with low energy shocks. Compared with low energy shocks, impedance measurement with subthreshold pulses has higher reproducibility and a higher correlation with the impedance obtained by high energy shock delivery. Safe and painless high energy impedance estimation with subthreshold pulses might, therefore, help to detect ICD lead failure during routine follow-up. (PACE 2003; 26:[Pt. II]:457,460) [source]


Safety of Pacemaker Implantation Prior to Radiofrequency Ablation of Atrioventricular Junction in a Single Session Procedure

PACING AND CLINICAL ELECTROPHYSIOLOGY, Issue 6 2000
ALESSANDRO PROCLEMER
RF current delivery may cause acute and chronic dysfunction of previously implanted pacemakers. The aim of this study was to assess prospectively the effects of RF energy on Thera I and Kappa pacemakers in 70 consecutive patients (mean age 70 ± 11 years, mean left ventricular ejection fraction 48 ± 15%) who underwent RF ablation of the AV junction for antiarrhythmic drug refractory atrial fibrillation (permanent in 42 patients, paroxysmal in 28). These pacing systems incorporate protection elements to avoid electromagnetic interference. The pacemakers (Thera DR 7960 I in 20 patients, Thera SR 8960 1 in 30, Kappa DR 600,601 in 8, Kappa SR 700,701 in 12) were implanted prior to RF ablation in a single session procedure and were transiently programmed to VVI mode at a rate of 30 beats/min. Capsure SP and Z unibipolar leads were used. During RF application there was continuous monitoring of three ECG leads, endocavitary electrograms, and event markers. Complete AV block was achieved in all cases after 3.6 ± 2.9 RF pulses and 100 ± 75 seconds of RF energy delivery. The mean time of pacemaker implantation and RF ablation was 60 ± 20 minutes. Transient or permanent pacemaker dysfunction including under/oversensing, reversion to a "noise-mode" pacing, pacing inhibition, reprogramming, or recycling were not observed. Leads impedance, sensing, and pacing thresholds remained in the normal range in the acute and long-term phase (average follow-up 18 ± 12 months). In conclusion, Thera I and Kappa pacemakers exhibit excellent protection against interference produced by RF current. The functional integrity of the pacemakers and Capsure leads was observed in the acute and chronic phases. Thus, the implantation of these pacing systems prior to RF ablation of the AV junction can be recommended. [source]


Intraoperative Comparison of a Subthreshold Test Pulse with the Standard High-Energy Shock Approach for the Measurement of Defibrillation Lead Impedance

JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, Issue 1 2006
ANDREAS SCHUCHERT M.D.
There are two methods to measure shocking lead impedance: delivery of high-energy shocks that require patient sedation, and the painless measurement of impedance from subthreshold test pulses. The aim of this study was to compare the two methods. Methods: The study included 131 patients implanted with a standard DR (n = 71) or VR (n = 60) ICD connected to either single-coil (n = 39) or dual-coil (n = 92) defibrillation leads. The noninvasive high-energy impedance test was done using a 17 J shock after induction of ventricular tachyarrhythmias and compared to a 0.4 ,J test pulse used by the ICD for the subthreshold measurements. Results: Defibrillation lead impedance measurements were not significantly different between patients with the same shocking vector configuration. In patients with a single-coil defibrillation lead the impedance was 62 ± 9 , with the high-energy shock and 62 ± 8 , with the subthreshold test pulses (P = 0.13). Patients with a dual-coil configuration recorded average impedances of 40 ± 5 , from both tests (P = 0.44). While there was no difference in values recorded within each lead configuration, there was a significant difference in impedance between the single-coil and the dual-coil patient groups (P = 0.001). Conclusions: There was no significant difference between shocking lead impedances measured with the high-energy shock or the subthreshold test pulses. This offers the possibility of noninvasive, low-energy serial measurements of shocking lead impedance at follow-up visits and removing the need for sedation. [source]


"Tuned" Defibrillation Waveforms Outperform 50/50% Tilt Defibrillation Waveforms: A Randomized Multi-Center Study

PACING AND CLINICAL ELECTROPHYSIOLOGY, Issue 2007
SENTHIL NATARAJAN M.D.
Introduction: A superior performance of a tuned waveform based on duration using an assumed cardiac membrane time constant of 3.5 ms and of a 50/50% tilt waveform over a standard 65/65% tilt waveform has been documented before. However, there has been no direct comparison of the tuned versus the 50/50% tilt waveforms. Methods: In 34 patients, defibrillation thresholds (DFTs) for tuned versus 50/50% tilt waveforms in a random order were measured by using the optimized binary search method. High voltage lead impedance was measured and used to select the pulse widths for tuned and 50/50% tilt defibrillation waveforms. Results: Delivered energy (7.3 ± 4.6 J vs 8.7 ± 5.3 J, P = 0.01), stored energy (8.2 ± 5.1 J vs 9.7 ± 5.6 J, P = 0.01), and delivered voltage (405.9 ± 121.7 V vs 445.0 ± 122.6 V, P = 0.008) were significantly lower for the tuned than for the 50/50% tilt waveform. In four patients with DFT ,15 J, the tuned waveform lowered the mean energy DFT by 2.8 J and mean voltage DFT by 45 V. For all patients, the mean peak delivered energy DFT was reduced from 29 J to 22 J (24% decrease). Multiple regression analysis showed that a left ventricular ejection fraction <20% is a significant predictor of this advantage. Conclusion: Energy and voltage DFTs are lowered with an implantable cardioverter defibrillator that uses a tuned waveform compared to a standard 50% tilt biphasic waveform. [source]


Subthreshold Test Pulses Versus Low Energy Shock Delivery to Estimate High Energy Lead Impedance in Implanted Cardioverter Defibrillator Patients

PACING AND CLINICAL ELECTROPHYSIOLOGY, Issue 1p2 2003
DIRK VOLLMANN
VOLLMANN, D., et al.: Subthreshold Test Pulses Versus Low Energy Shock Delivery to Estimate High Energy Lead Impedance in Implanted Cardioverter Defibrillator Patients. The high energy lead impedance is valuable for detecting lead failure in ICDs, but until recently shock delivery was necessary for high energy impedance measurement. This study compared the use of subthreshold test pulses and low energy test shocks to estimate the high energy impedance. Immediately after implantation of Ventak Prizm ICDs in 29 patients, the lead impedance was measured with five subthreshold (0.4 ,J) test pulses, 5 low energy (1.1 J) shocks, and two to three high energy(16 ± 4.5 J)shocks. The mean impedances measured using high energy shocks, low energy shocks, and subthreshold pulses were42.0 ± 7.3 ,, 46.5 ± 8.1 ,, and42.4 ± 7.1 ,, respectively. The impedances measured using high and low energy shocks differed significantly(P <0.0001), while those obtained by high energy shocks and low energy pulses did not(P = 0.63). According to the Pearson correlation coefficient, the impedance measurements with subthreshold pulses and low energy shocks were both closely correlated(P < 0.0001)with impedance values determined with high energy shocks. However, while the impedance values tended to be higher when measured with low energy shocks, the concordance correlation coefficient (c) was higher for subthreshold test pulse versus high energy shock(c = 0.92)than for low versus high energy shock(c = 0.73). Furthermore, the intraindividual variability of impedance measurements was lower with subthreshold pulse measurements than with low energy shocks. Compared with low energy shocks, impedance measurement with subthreshold pulses has higher reproducibility and a higher correlation with the impedance obtained by high energy shock delivery. Safe and painless high energy impedance estimation with subthreshold pulses might, therefore, help to detect ICD lead failure during routine follow-up. (PACE 2003; 26:[Pt. II]:457,460) [source]


Compatibility of Automatic Threshold Tracking Pacemakers with Previously Implanted Pacing Leads in Children

PACING AND CLINICAL ELECTROPHYSIOLOGY, Issue 11 2002
OSMAN KUCUKOSMANOGLU
KUCUKOSMANOGLU, O., et al.: Compatibility of Automatic Threshold Tracking Pacemakers with Previously Implanted Pacing Leads in Children. The Autocapture function controls and optimizes the amplitude of the pacing pulse and saves energy. The manufacturer recommends using a special low polarization, low threshold bipolar Pacesetter lead for the Autocapture function. The purpose of this study was to evaluate the compatibility of Autocapture with previously implanted pacing leads. The study included 15 patients (mean age 13.6 ± 3.4 years) who needed pulse generator replacement and received the VVIR pacemaker Regency SR+ or the DDDR pacemakers Affinity DR or Integrity DR with the Autocapture function. The new pulse generators connected to previously implanted ventricular leads. At the time of implantation the pacing threshold was 1.0 ± 0.35 V at 0.5 ms, the lead impedance was 580 ± 80 ,, and the spontaneous R wave amplitude was 7.89 ± 4.89 mV. The polarization signal (PS) was 3.8 ± 3.04 mV, and evoked response (ER) was 8.15 ± 4.57 mV at the predischarge testing. Follow-up telemetry was done at months 1, 3, 6, 12, and 18. The follow-up duration was 9.4 ± 5 months (range 1,18 months). If the results of PS and ER measurements were acceptable for Autocapture, it turned on at the 1-month visit. In six (40%) patients the results were found acceptable for Autocapture function. Age, lead impedance, pacing threshold, intrinsic R wave measurement, lead age, fixation mechanism, and ER measurements were not statistically different in Autocapture suitable and not suitable groups. The main reason not to activate Autocapture had been increased PS. Any significant fluctuations were not observed in pacing threshold, lead impedance, ER, and PS during follow-up. In conclusion, previously implanted pacing leads may be compatible with the Autocapture function. [source]


Unexpected Loss of Bipolar Pacing With Implanted Dual Chamber Pacemakers

PACING AND CLINICAL ELECTROPHYSIOLOGY, Issue 1 2000
G. BIH-FANG GUO
Bipolar leads are most commonly used in the current practice of pacemaker therapy. In our study of 124 patients implanted with Guidant/Cardiac Pacemakers (CPI) Vigor dual chamber pacemakers, 5 patients had unexpectedly abrupt increases in bipolar lead impedance and pacing threshold 2 weeks to 18 months postimplantation without changes in sensing function. With the lead configuration reprogrammed to unipolar, the lead impedance and pacing threshold were restored to appropriate ranges. The changes in bipolar lead parameters can be caused by the CPI's "Quick Connect" (QC1) header lead system incorporated in these pacemakers. [source]


Intraoperative Comparison of a Subthreshold Test Pulse with the Standard High-Energy Shock Approach for the Measurement of Defibrillation Lead Impedance

JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, Issue 1 2006
ANDREAS SCHUCHERT M.D.
There are two methods to measure shocking lead impedance: delivery of high-energy shocks that require patient sedation, and the painless measurement of impedance from subthreshold test pulses. The aim of this study was to compare the two methods. Methods: The study included 131 patients implanted with a standard DR (n = 71) or VR (n = 60) ICD connected to either single-coil (n = 39) or dual-coil (n = 92) defibrillation leads. The noninvasive high-energy impedance test was done using a 17 J shock after induction of ventricular tachyarrhythmias and compared to a 0.4 ,J test pulse used by the ICD for the subthreshold measurements. Results: Defibrillation lead impedance measurements were not significantly different between patients with the same shocking vector configuration. In patients with a single-coil defibrillation lead the impedance was 62 ± 9 , with the high-energy shock and 62 ± 8 , with the subthreshold test pulses (P = 0.13). Patients with a dual-coil configuration recorded average impedances of 40 ± 5 , from both tests (P = 0.44). While there was no difference in values recorded within each lead configuration, there was a significant difference in impedance between the single-coil and the dual-coil patient groups (P = 0.001). Conclusions: There was no significant difference between shocking lead impedances measured with the high-energy shock or the subthreshold test pulses. This offers the possibility of noninvasive, low-energy serial measurements of shocking lead impedance at follow-up visits and removing the need for sedation. [source]